Penetrating microelectrode arrays with low-impedance sputtered iridium oxide electrode coatings
Authors:
- 1 September 2009
- conference paper
- Published by Institute of Electrical and Electronics Engineers (IEEE)
- Vol. 2009 (1094687X) , 7147-7150
Abstract
Sputtered iridium oxide (SIROF) is a candidate low-impedance coating for neural stimulation and recording electrodes. SIROF on planar substrates has exhibited a high charge-injection capacity and impedance suitable for indwelling cortical microelectrode applications. In the present work, the properties of SIROF electrode coatings deposited onto multi-shank penetrating arrays intended for intracortical and intraneural applications were examined. The charge-injection properties under constant current pulsing were evaluated for a range of pulsewidths and current densities using voltage transients to determine maximum potential excursions in an inorganic model of interstitial fluid at 37degC. The charge-injection capacity of the SIROFs was significantly improved by the use of positive potential biasing in the interpulse period, but even without bias, the SIROFs reversibly inject higher charge than other iridium oxides or platinum. Typical deliverable charge levels of 25 to 160 nC/phase were obtained with 2000 mum2 electrodes depending on pulsewidth and interpulse bias. Similar sized platinum electrodes could inject 3 to 8 nC/phase.Keywords
This publication has 13 references indexed in Scilit:
- Flexible Nerve Stimulation Electrode With Iridium Oxide Sputtered on Liquid Crystal PolymerIEEE Transactions on Biomedical Engineering, 2009
- Sputtered iridium oxide films for neural stimulation electrodesJournal of Biomedical Materials Research Part B: Applied Biomaterials, 2008
- Neural Stimulation and Recording ElectrodesAnnual Review of Biomedical Engineering, 2008
- The influence of electrolyte composition on thein vitrocharge-injection limits of activated iridium oxide (AIROF) stimulation electrodesJournal of Neural Engineering, 2007
- "Safe" charge-injection waveforms for iridium oxide (AIROF) microelectrodes2007
- Neurotrophin‐eluting hydrogel coatings for neural stimulating electrodesJournal of Biomedical Materials Research Part B: Applied Biomaterials, 2006
- Potential-Biased, Asymmetric Waveforms for Charge-Injection With Activated Iridium Oxide (AIROF) Neural Stimulation ElectrodesIEEE Transactions on Biomedical Engineering, 2006
- In Vitro Comparison of the Charge-Injection Limits of Activated Iridium Oxide (AIROF) and Platinum-Iridium MicroelectrodesIEEE Transactions on Biomedical Engineering, 2005
- Electrical stimulation of excitable tissue: design of efficacious and safe protocolsJournal of Neuroscience Methods, 2005
- The Effects of Prolonged Intracortical Microstimulation on the Excitability of Pyramidal Tract Neurons in the CatAnnals of Biomedical Engineering, 2002