Influence of macromolecule baseline on 1H MR spectroscopic imaging reproducibility
Open Access
- 22 January 2016
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 77 (1) , 34-43
Abstract
Purpose: Poorly characterized macromolecular (MM) and baseline artefacts are known to reduce metabolite quantitation accuracy in 1H MR spectroscopic imaging (MRSI). Increasing echo time (TE) and improvements in MM analysis schemes have both been proposed as strategies to improve metabolite measurement reliability. In this study, the influence of TE and two MM analysis schemes on MRSI reproducibility are investigated.Methods: An experimentally acquired baseline was collected using an inversion recovery sequence (TI = 750 ms) and incorporated into the analysis method. Intrasubject reproducibility of MRSI scans, acquired at 3 Tesla, was assessed using metabolite coefficients of variance (COVs) for both experimentally acquired and simulated MM analysis schemes. In addition, the reproducibility of TE = 35 ms, 80 ms, and 144 ms was evaluated.Results: TE = 80 ms was the most reproducible for singlet metabolites with COVs < 6% for total N‐acetyl‐aspartate, total creatine, and total choline; however, moderate multiplet dephasing was observed. Analysis incorporating the experimental baseline achieved higher Glu and Glx reproducibility at TE = 35 ms, and showed improvements over the simulated baseline, with higher efficacy for poorer data.Conclusion: Overall, TE = 80 ms yielded the most reproducible singlet metabolite estimates. However, combined use of a short TE sequence and the experimental baseline may be preferred as a compromise between accuracy, multiplet dephasing, and T2 bias on metabolite estimates. Magn Reson Med 77:34–43, 2017. © 2016 The Authors Magnetic Resonance in Medicine published by Wiley Periodicals, Inc. on behalf of International Society for Magnetic Resonance in Medicine.Keywords
Funding Information
- Engineering and Physical Sciences Research Council
- National Institute for Health Research (13‐0053)
This publication has 45 references indexed in Scilit:
- Magnetic resonance spectroscopy metabolite profiles predict survival in paediatric brain tumoursEuropean Journal of Cancer, 2013
- Reproducibility of rapid short echo time CSI at 3 tesla for clinical applicationsJournal of Magnetic Resonance Imaging, 2012
- Test‐retest reliability and reproducibility of short‐echo‐time spectroscopic imaging of human brain at 3TMagnetic Resonance in Medicine, 2011
- MR spectroscopy-based brain metabolite profiling in propionic acidaemia: metabolic changes in the basal ganglia during acute decompensation and effect of liver transplantationOrphanet Journal of Rare Diseases, 2011
- Reproducibility of serial whole‐brain MR Spectroscopic ImagingNMR in Biomedicine, 2010
- Optimized glutamate detection at 3TJournal of Magnetic Resonance Imaging, 2009
- Comparative reliability of proton spectroscopy techniques designed to improve detection of J‐coupled metabolitesMagnetic Resonance in Medicine, 2008
- In vivo proton MR spectroscopy of the human brainProgress in Nuclear Magnetic Resonance Spectroscopy, 2006
- Noninvasive Magnetic Resonance Spectroscopic Imaging Biomarkers to Predict the Clinical Grade of Pediatric Brain TumorsClinical Cancer Research, 2004
- 1H MR spectroscopy of brain tumours and massesNMR in Biomedicine, 2003