CIED malfunction in patients receiving radiation is a rare event that could be detected by remote monitoring
- 4 June 2018
- journal article
- research article
- Published by Wiley in Journal of Cardiovascular Electrophysiology
- Vol. 29 (9) , 1268-1275
Abstract
Introduction: An increasing number of patients with cardiac devices require radiation therapy for treatment of a variety of cancers. This study aimed to identify the incidence and predictors of cardiac implantable electronic devices (CIED) malfunction in a real‐world population that has received radiation therapy.Methods: This retrospective cohort study included 109 adult patients who received radiation therapy at the University of Rochester Medical Center, Radiation Oncology Department, between 2000 and 2015. Sixty patients had pacemakers and 49 had automatic implantable cardioverter defibrillators. Subjects received either high energy (16 MV) and/or low energy (6 MV) photon beams with or without electron beams (6–16 MeV). We included interrogations done from first day of radiation and up to 3 months’ postradiation therapy. Outcomes analyzed were device‐related malfunctions and device‐related clinical events. Fisher's exact, Wilcoxon, and Kruskall‐Wallis tests were used for bivariate analysis. Logistic regression with robust adjustment was used for multivariate analysis.Results: We identified six device‐related malfunctions. All events were minor and included partial settings reset leading to loss of historical data, pacing thresholds changes, lead impedance changes, and LV output increase. Two patients had device‐related clinical events, including dyspnea and diaphragmatic‐stimulation. In bivariate analysis, CIED malfunction was associated with CIED duration in situ. In multivariate analysis, there was no significant statistical association between adverse events and beam energy type, CIED location, or dose of radiation delivered to the target.Conclusions: CIED malfunctions are uncommon in real‐world patients and associated with minor clinical events. In our cohort, remote CIED monitoring would have identified all events.Keywords
Funding Information
- CTSI University of Rochester (3657)
This publication has 17 references indexed in Scilit:
- A review of implantable cardioverter defibrillator failures during radiation therapy in three Sydney hospitalsJournal of Medical Imaging and Radiation Oncology, 2017
- Radiotherapy-Induced Cardiac Implantable Electronic Device Dysfunction in Patients With CancerThe American Journal of Cardiology, 2017
- Impact of remote monitoring on clinical events and associated health care utilization: A nationwide assessmentHeart Rhythm, 2016
- Management of patients with implantable cardioverter-defibrillators and pacemakers who require radiation therapyHeart Rhythm, 2015
- Radiotherapy-Induced Malfunction in Contemporary Cardiovascular Implantable Electronic DevicesJAMA Oncology, 2015
- Effect of radiation therapy on permanent pacemaker and implantable cardioverter-defibrillator functionHeart Rhythm, 2012
- High incidence of implantable cardioverter defibrillator malfunctions during radiation therapy: neutrons as a probable cause of soft errorsEP Europace, 2012
- The Heart Rhythm Society (HRS)/American Society of Anesthesiologists (ASA) Expert Consensus Statement on the Perioperative Management of Patients with Implantable Defibrillators, Pacemakers and Arrhythmia Monitors: Facilities and Patient ManagementHeart Rhythm, 2011
- Implanted Cardiac Defibrillator Care in Radiation Oncology Patient PopulationInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2009
- Treatment of patients with cardiac pacemakers and implantable cardioverter-defibrillators during radiotherapyInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2004