Assessment of lower tear meniscus measurements obtained with Keratograph and agreement with Fourier-domain optical-coherence tomography
- 12 February 2015
- journal article
- research article
- Published by BMJ in British Journal of Ophthalmology
- Vol. 99 (8) , 1120-1125
Abstract
Aims To compare intraobserver repeatability and interobserver and intersession reproducibility of lower tear meniscus height measurements (LTMH) obtained with the new Keratograph 5M and a Fourier-domain optical-coherence tomography (OCT; Spectralis OCT) and to assess the agreement between the two devices. Methods This is a observational cross-sectional study. Thirty eyes of 30 normal subjects were randomly imaged twice with both devices in our setting. Two examiners randomly made LTMH measurements using the devices. To assess intraobserver repeatability and interobserver and intersession reproducibility within-subject SD (Sw), test–retest repeatability, coefficient of variation (CoV) and intraclass correlation coefficients (ICCs) were calculated. Agreement between both devices was also determined. Results Average LMTH for Keratograph was 235.46±57.61 μm, whereas the mean LTMH measured by OCT was 245.84±67.18 μm. Intraobserver repeatability and interobserver and intersession reproducibility were moderate with Keratograph (CoV ≥0.16%, ≥0.19%, ≥0.17% and ICC ≤0.83, ≤0.76, ≤0.83, respectively) and high for the Spectralis OCT (CoV ≤0.13%, ≤0.12%, ≤0.11% and ICC ≥0.9, ≥0.92, ≥0.92, respectively). No statistically significant difference was detected in mean LMTH measurements made with both devices, although correlation was low (CoV 0.25%, ICC 0.55). Conclusions Spectralis OCT LTMH measurements were more reliable than the new Keratograph data. Agreement between the devices was poor.Keywords
This publication has 24 references indexed in Scilit:
- Correlation Between Optical Coherence Tomography–Derived Assessments of Lower Tear Meniscus Parameters and Clinical Features of Dry Eye DiseaseCornea, 2012
- Between-Grader Repeatability of Tear Meniscus Measurements Using Fourier-Domain OCT in Patients With Dry EyeOphthalmic Surgery, Lasers and Imaging Retina, 2011
- Application of Visante Optical Coherence Tomography Tear Meniscus Height Measurement in the Diagnosis of Dry Eye DiseaseOphthalmology, 2010
- Reproducibility of Tear Meniscus Measurement by Fourier-Domain Optical Coherence Tomography: A Pilot StudyOphthalmic Surgery, Lasers and Imaging Retina, 2009
- Upper and Lower Tear Menisci in the Diagnosis of Dry EyeInvestigative Ophthalmology & Visual Science, 2009
- Variability of the Analysis of the Tear Meniscus Height by Optical Coherence TomographyOptometry and Vision Science, 2007
- The Epidemiology of Dry Eye Disease: Report of the Epidemiology Subcommittee of the International Dry Eye WorkShop (2007)The Ocular Surface, 2007
- The Definition and Classification of Dry Eye Disease: Report of the Definition and Classification Subcommittee of the International Dry Eye Workshop (2007)The Ocular Surface, 2007
- The Repeatability of Clinical Measurements of Dry EyeCornea, 2004
- The Height and Radius of the Tear Meniscus and Methods for Examining These ParametersCornea, 2000