Utilizing a Functional Lumen Imaging Probe for Evaluation of the Pediatric Airway, A Pilot Study
Open Access
- 17 May 2023
- journal article
- research article
- Published by Wiley in The Laryngoscope
- Vol. 134 (1) , 108-112
Abstract
Objectives: Accurate and reproducible measurements of the pediatric airway are critical for diagnostic evaluation and management of subglottic and tracheal stenosis. The endoluminal functional lumen imaging probe (EndoFLIP) is a catheter‐based imaging probe which utilizes impedance planimetry to calculate luminal parameters, including cross‐sectional area and compliance. Herein, we demonstrate the feasibility of this system for multidimensional evaluation of the pediatric airway.Methods: 3D‐printed pediatric laryngotracheal models were created based on computed tomography scans, then artificially deformed to simulate both circumferential and posterior subglottic stenosis. Two observers made six measurements of the minimum cross‐sectional area (MCSA) and length of stenosis of each model with EndoFLIP. Agreement between observer measurements and model dimensions was evaluated using Lin's concordance correlation coefficient; inter‐observer reliability was assessed using intraclass correlation.Results: Four models were created: two without pathology (MCSA: 132.4, 44.3 mm2) and two with subglottic stenosis (MCSA: 28.7, 59.7 mm2, stenotic length 27.8, 24.4 mm). Observer measurements of MCSA and length of stenosis demonstrated high concordance with the models (r = 0.99, 0.95, p < 0.001) with a mean error of 4.5% and 18.2% respectively. There was a low coefficient of variation (0.6%–2.8%) for measurements, indicating high precision. Interrater reliability was high for both MCSA and stenotic length (ICC: 0.99, 0.98).Conclusions: The EndoFLIP system allows for accurate and reproducible measurements of cross‐sectional area and stenotic length in pediatric airway models. This method may provide further advantages in the evaluation of airway distensibility, as well as measurements of asymmetric airway pathology.Level of Evidence: NA Laryngoscope, 134:108–112, 2024Keywords
Funding Information
- National Institute on Deafness and Other Communication Disorders (5T32DC000027)
This publication has 19 references indexed in Scilit:
- Mechanical Modeling of the Human Cricoid Cartilage Using Computer-Aided DesignAnnals of Otology, Rhinology & Laryngology, 2015
- A Novel Device for Measurement of Subglottic Stenosis in 3 Dimensions During Suspension LaryngoscopyJAMA Otolaryngology–Head & Neck Surgery, 2015
- Cervical Stiffness Evaluated In Vivo by Endoflip in Pregnant WomenPLOS ONE, 2014
- Novel Measurements of the Length of the Subglottic Airway in Infants and Young ChildrenAnesthesia & Analgesia, 2013
- Anatomical reconstructions of pediatric airways from endoscopic images: A pilot study of the accuracy of quantitative endoscopyThe Laryngoscope, 2013
- Incidence of post-intubation subglottic stenosis in children: prospective studyThe Journal of Laryngology & Otology, 2013
- Outcomes of Balloon Dilation in Pediatric Subglottic StenosisAnnals of Otology, Rhinology & Laryngology, 2012
- Anatomical Reconstruction from Endoscopic Images: Toward Quantitative EndoscopyAmerican Journal of Rhinology, 2008
- Proposed Grading System for Subglottic Stenosis Based on Endotracheal Tube SizesAnnals of Otology, Rhinology & Laryngology, 1994
- Acquired and Congenital Subglottic Stenosis in the InfantAnnals of Otology, Rhinology & Laryngology, 1976