A PET Design Based on SiPM and Monolithic LYSO Crystals: Performance Evaluation
- 20 June 2016
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Nuclear Science
- Vol. 63 (5) , 2471-2477
Abstract
A new small animal PET based on SiPM and monolithic LYSO crystals has been developed. Eight detector modules form the PET ring, each mounting an array of 12 × 12 SiPMs coupled to a readout providing the summed signals of the pixels on each of the 12 rows and 12 columns of the SiPM array. This design makes it possible to accurately determine the centroid of the scintillation light distribution with about 1.6 mm full width at half maximum (FWHM) resolution without correction for the 1 mm source size, and the photon depth of interaction (DOI) with nearly 2 mm FWHM. This single ring PET system has a homogeneous spatial resolution across the entire 80 mm transaxial field of view (FOV) of about 1 mm FWHM. The noise equivalent count rate (NECR) peak is estimated to occur at around 39.2 MBq with a rate of approximately 82.7 kcps for the mouse-like phantom and 22 kcps at 48.1 MBq for the rat-like phantom. Following the NEMA protocol, the peak absolute sensitivity in the center of the FOV is 2.8% for a 30% peak energy window. A pilot test injecting NaF to a mouse of 20 grams is also presented. Finally, the PET ring has been tested in front of a high field 15.2 T Magnetic Resonance (MR). No significant variation on energy and spatial resolution across the FOV has been observed due to the presence of the magnetic field.Keywords
Funding Information
- Spanish Plan Nacional de Investigación Científica, Desarrollo e Innovación Tecnológica (TEC 2013-48036-C3-1-R)
- Valencian Local Government (PROMETEOII/2013/010, ISIC 2011/013)
- CSIC with a JAE-Doc
- Junta de Ampliación de Estudios
- European Social Fund Plus
This publication has 23 references indexed in Scilit:
- 3-D photon impact determination using fitting approaches to the Light Distribution2014
- High-volume silicon photomultiplier production, performance, and reliabilityOptical Engineering, 2014
- Evaluation of Imaging Modules Based on SensL Array SB-8 for Nuclear Medicine ApplicationsIEEE Transactions on Nuclear Science, 2014
- Quantitative PET and SPECT performance characteristics of the Albira Trimodal pre-clinical tomographPhysics in Medicine & Biology, 2014
- ALBIRA: A small animal PET/SPECT/CT imaging systemMedical Physics, 2013
- Small animal PET scanner based on monolithic LYSO crystals: Performance evaluationMedical Physics, 2012
- Continuous DoI determination by gaussian modelling of linear and non-linear scintillation light distributions2011
- Silicon Photomultipliers (SiPM) as novel photodetectors for PETNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2011
- Design and Calibration of a Small Animal Pet Scanner Based on Continuous LYSO Crystals and PSPMTs2006
- Design and tests of a portable mini gamma cameraMedical Physics, 2004