Quality by design tool–evaluated stability‐indicating ultra‐performance liquid chromatography method for the determination of drugs (ritonavir and darunavir) used to treat the human immunodeficiency virus/acquired immunodeficiency syndrome
- 1 July 2023
- journal article
- research article
- Published by Wiley in Biomedical Chromatography
- Vol. 37 (9) , e5687
Abstract
Ritonavir and darunavir were examined using a ultra‐performance liquid chromatography (UPLC) approach in pharmaceutical dosage forms. The small number of analytical studies that are currently available do not demonstrate the method's stability or nature. The study sought to assess both chemicals using a stability‐indicating approach with a relatively short run time. The HSS C18 (100 × 2.1 mm), 2‐mm column was used for the chromatographic separation, and isocratic elution was used to achieve this. In the mobile phase, methanol and 0.01 M phosphate buffer (pH 4.0) were included in a 60:40 (v/v) ratio. Throughout the analysis, the flow rate was kept at 0.2 mL min−1, and a photodiode array detector set to 266 nm was used to find the major components. The proposed method showed a linear response (r2 > 0.999), and the accuracy was between 98.0% and 102.0%. The precision data showed relative standard deviation ≤1.0%. The UPLC method for quantification of ritonavir and darunavir in pharmaceutical dosage forms using a very short run time of under a minute is the subject of the proposed article. To meet current regulatory criteria, the quality by design idea was used in the method performance verification.Keywords
This publication has 20 references indexed in Scilit:
- Analytical method development and validation for the simultaneous estimation of Darunavir and Ritonavir by RP-HPLC methodThe World Journal of Pharmaceutical Sciences, 2022
- Eco-friendly and green chromatographic method for the simultaneous determination of chlorocresol and betamethasone dipropionate in topical formulations using Box–Behnken designJournal of the Iranian Chemical Society, 2021
- Development and Validation of New Analytical Method for The Simultaneous Estimation of Darunavir And Ritonavir in Pharmaceutical Dosage FormInternational Journal of Indigenous Herbs and Drugs, 2021
- Development and validation of simplified RP-HPLC method for quantification of Darunavir in commercial tabletsMaterials Today: Proceedings, 2021
- A Novel Ultra Performance Liquid Chromatography-PDA Method Development and Validation for Darunavir in Bulk and Its Application to Marketed Dosage FormJournal of Pharmacy and Bioallied Sciences, 2020
- Analytical Techniques in the Analysis of Darunavir and Ritonavir: A ReviewCurrent Pharmaceutical Analysis, 2020
- Physiologically Based Modelling of Darunavir/Ritonavir Pharmacokinetics During PregnancyClinical Pharmacokinetics, 2015
- Validation of Simultaneous Quantitative Method of HIV Protease Inhibitors Atazanavir, Darunavir and Ritonavir in Human Plasma by UPLC-MS/MSThe Scientific World Journal, 2014
- A novel LC–ESI-MS method for the simultaneous determination of etravirine, darunavir and ritonavir in human blood plasmaTalanta, 2009
- The Validation of Plasma Darunavir Concentrations Determined by the HPLC Method for Protease InhibitorsBiological and Pharmaceutical Bulletin, 2007