Performance of the HIV Blot 2.2, INNO-LIA HIV I/II Score, and Geenius HIV 1/2 Confirmatory Assay for use in HIV confirmation
Open Access
- 21 June 2018
- journal article
- research article
- Published by Public Library of Science (PLoS) in PLOS ONE
- Vol. 13 (6) , e0199502
Abstract
In view of recent revised recommendations for human immunodeficiency virus (HIV) confirmatory testing, the performance of 3 HIV confirmatory assays was compared. Using the HIV Blot 2.2 (MP-WB), the INNO-LIA HIV I/II Score (INNO), and the Geenius HIV 1/2 Confirmatory Assay (Geenius), we tested 199 HIV-1 positive, 161 HIV negative, 65 HIV western blot indeterminate, 26 HIV seroconversion, 34 early HIV infection and 4 HIV-2 positive archived specimens. We show that all 3 assays had comparable test sensitivity in the detection of HIV-1 positive cases. However, less non-specific reactivity was observed with the INNO and Geenius assays, where both of them were able to resolve MP-WB indeterminate cases. When early HIV cases were considered, INNO and Geenius were more likely to confirm an early-stage infection as positive. Nevertheless, overall poor sensitivity (25.5% - 44.7%) of these assays for the detection of early cases was observed, likely because these cases had very low or non-detectable levels of HIV antibodies. Hence, further testing by a nucleic acid test or a p24 antigen test of specimens reactive on screening with a fourth generation Ag/Ab assay that are negative on confirmatory testing for HIV-specific antibody, may be useful. In conclusion, INNO and Geenius had comparable test performance, although the ease of use and shorter assay time for Geenius may make it the preferred choice for laboratories. In that regard, of note is our observation of non-specific reactivity of lipaemic specimens to the HIV-2 gp140 band in the Geenius assay, which should prompt caution when interpreting results of such specimens.Keywords
This publication has 15 references indexed in Scilit:
- Comparison of a conventional HIV 1/2 line immunoassay with a rapid confirmatory HIV 1/2 assayJournal of Virological Methods, 2014
- Evaluation of the Bio-Rad Geenius HIV-1/2 test as a confirmatory assayJournal of Clinical Virology, 2014
- Field evaluation of the Abbott ARCHITECT HIV Ag/Ab Combo immunoassayJournal of Clinical Virology, 2013
- Identification of early HIV infections using the fourth generation Abbott ARCHITECT HIV Ag/Ab Combo chemiluminescent microparticle immunoassay (CIA) in San Diego CountyJournal of Clinical Virology, 2013
- Clinical performance of the Multispot HIV-1/HIV-2 rapid test to correctly differentiate HIV-2 from HIV-1 infection in screening algorithms using third and fourth generation assays and to identify cross reactivity with the HIV-1 Western BlotJournal of Clinical Virology, 2013
- Special Aspects of the Treatment of HIV-2-Infected PatientsIntervirology, 2012
- Comparison of alternative interpretive criteria for the HIV-1 Western blot and results of the Multispot HIV-1/HIV-2 Rapid Test for classifying HIV-1 and HIV-2 infectionsJournal of Clinical Virology, 2011
- HIV Type 2 in New York City, 2000–2008Clinical Infectious Diseases, 2010
- The Measurement of Interrater Agreement2003
- A Coefficient of Agreement for Nominal ScalesEducational and Psychological Measurement, 1960