Endovascular coil detachment causing EMG artefact in BIS: a mechanistic exploration
- 7 May 2015
- journal article
- case report
- Published by Springer Nature in Journal of Clinical Monitoring and Computing
- Vol. 30 (2) , 203-206
Abstract
Deployment of endovascular coils used in interventional neuroradiology commonly involves electrolytic detachment of the coil from the pusher catheter. This report describes a case of artefactual increase in electromyography (EMG) values of bispectral index (BIS) monitor during coil detachment. An explanation of this event is provided connecting mechanism of coil detachment and derivation of EMG values in a BIS monitor. While rising EMG values are thought to arise from frontalis contraction, they may as well be an unrecognized electrical artefact, especially in context of undistorted electroencephalography waveform.Keywords
This publication has 3 references indexed in Scilit:
- Diagnostic and prognostic use of bispectral index in coma, vegetative state and related disordersBrain Injury, 2008
- EMG contamination of EEG: spectral and topographical characteristicsClinical Neurophysiology, 2003
- A Primer for EEG Signal Processing in AnesthesiaAnesthesiology, 1998