Multi-contrast CT imaging using a high energy resolution CdTe detector and a CZT photon-counting detector
Authors:
- 1 January 2022
- journal article
- research article
- Published by IOP Publishing in Journal of Instrumentation
- Vol. 17 (01) , P01004
Abstract
Computed tomography (CT) imaging with high energy resolution detectors shows great promise in material decomposition and multi-contrast imaging. Multi-contrast imaging was studied by imaging a phantom with iodine (I), gadolinium (Gd), and gold (Au) solutions, and mixtures of the three using a cadmium telluride (CdTe) spectrometer with an energy resolution of 1% as well as with a cadmium zinc telluride (CZT) detector with an energy resolution of 13%. The phantom was imaged at 120 kVp and 1.1 mA with 7 mm of aluminum filtration. For the CdTe data collection, the phantom was imaged using a 0.2 mm diameter x-ray beam with 96 ten-second data acquisitions across the phantom at 45 rotation angles. For the CZT detector, we had 720 projections using a cone beam, and the six detector energy thresholds were set to 23, 33, 50, 64, 81, and 120 keV so that three thresholds corresponded to the K-edges of the contrast agents. Contrast agent isolation methods were then examined. K-edge subtraction and novel spectrometric algebraic image reconstruction (SAIR) were used for the CdTe data. K-edge subtraction alone was used for the CZT data. Linearity plots produced similar R 2 values and slopes for all three reconstruction methods. Comparing CdTe methods, SAIR offered less noise than CdTe K-edge subtraction and better geometric accuracy at low contrast concentrations. CdTe contrast agent images of I, Gd, and Au offered less noise and greater contrast than the CZT images, highlighting the benefits of high energy resolution CdTe detectors for possible use in pre-clinical or clinical CT imaging.Keywords
This publication has 26 references indexed in Scilit:
- Vascular imaging with contrast agent in hard and soft tissues using microcomputed‐tomographyJournal of Microscopy, 2015
- Dual- and Multi-Energy CT: Principles, Technical Approaches, and Clinical ApplicationsRadiology, 2015
- Soft tissue imaging with photon counting spectroscopic CTPhysics in Medicine & Biology, 2015
- Blood vessel‐based liver segmentation using the portal phase of an abdominal CT datasetMedical Physics, 2013
- An exploratory study of contrast agents for soft tissue visualization by means of high resolution X‐ray computed tomography imagingJournal of Microscopy, 2013
- TOPAS: An innovative proton Monte Carlo platform for research and clinical applicationsMedical Physics, 2012
- Performance of a digital CdTe X-ray spectrometer in low and high counting rate environmentNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2010
- Monte Carlo simulation of a computed tomography x-ray tubePhysics in Medicine & Biology, 2007
- Practical cone-beam algorithmJournal of the Optical Society of America A, 1984
- Computerized transverse axial scanning (tomography): Part 1. Description of systemBritish Journal of Radiology, 1973