Pulse oximeter bench tests under different simulated skin tones
Open Access
- 24 April 2024
- journal article
- research article
- Published by Springer Nature in Medical & Biological Engineering & Computing
- Vol. 63 (7) , 1931-1942
Abstract
Pulse oximeters’ (POs) varying performance based on skin tones has been highly publicised. Compared to arterial blood gas analysis, POs tend to overestimate oxygen saturation (SpO 2 ) values for people with darker skin (occult hypoxemia). The objective is to develop a test bench for assessing commercial home and hospital-based POs in controlled laboratory conditions. A laboratory simulator was used to mimic different SpO 2 values (~ 70 to 100%). Different neutral density and synthetic melanin filters were used to reproduce low signal and varying melanin attenuation levels. Six devices consisting of commercial home (Biolight, N = 13; ChoiceMMed, N = 18; MedLinket, N = 9) and hospital-based (Masimo Radical 7 with Neo L, N = 1; GE B450 Masimo SET with LNCS Neo L, N = 1; Nonin 9550 Onyx II™, N = 1) POs were reviewed and their response documented. Significant variations were observed in the recorded SpO 2 values among different POs when exposed to identical simulated signals. Differences were greatest for lower SpO 2 (< 80%) where empirical data is limited. All PO responses under low signal and melanin attenuation did not change across various simulated SpO 2 values. The bench tests do not provide conclusive evidence that melanin does not affect in vivo SpO 2 measurements. Research in the areas of instrument calibration, theory and design needs to be further developed. Graphical AbstractKeywords
Funding Information
- Engineering and Physical Sciences Research Council (EP/T517902/1)
- Medical Research Council & NHS@home programme (MR/T025638/1)
This publication has 22 references indexed in Scilit:
- Factors associated with enrolment into a national COVID-19 pulse oximetry remote monitoring programme in England: a retrospective observational studyThe Lancet Digital Health, 2023
- Differential pulse oximetry readings between ethnic groups and delayed transfer to intensive care unitsQJM: An International Journal of Medicine, 2022
- Assessment of Racial and Ethnic Differences in Oxygen Supplementation Among Patients in the Intensive Care UnitJAMA Internal Medicine, 2022
- Racial and Ethnic Discrepancy in Pulse Oximetry and Delayed Identification of Treatment Eligibility Among Patients With COVID-19JAMA Internal Medicine, 2022
- Silent hypoxia in COVID-19: pathomechanism and possible management strategyMolecular Biology Reports, 2021
- Racial Bias in Pulse Oximetry MeasurementNew England Journal of Medicine, 2020
- Research Techniques Made Simple: Cutaneous Colorimetry: A Reliable Technique for Objective Skin Color MeasurementJournal of Investigative Dermatology, 2020
- Pulse oximetry: Understanding its basic principles facilitates appreciation of its limitationsRespiratory Medicine, 2013
- Photoplethysmography and its application in clinical physiological measurementPhysiological Measurement, 2007
- Non-invasive optical monitoring of the newborn piglet brain using continuous-wave and frequency-domain spectroscopyPhysics in Medicine & Biology, 1999