Imaging efficiency optimization for x-ray fluorescence computed tomography based on Thomson scattering x-ray sources
- 24 March 2026
- journal article
- Published by AIP Publishing in Journal of Applied Physics
- Vol. 139 (12)
Abstract
X-ray fluorescence computed tomography (XFCT) based on Thomson scattering x-ray sources is a novel molecular imaging technique applicable in both laboratory and clinical settings. However, its practical development is limited by relatively low imaging efficiency. To address this challenge, this study presents a dual approach that simultaneously optimizes the incident x-ray energy to maximize fluorescence yield and designs a three-pinhole collimator to enhance detection sensitivity. For the element gold (Au) widely used in XFCT imaging, the optimal incident x-ray energy was determined to be 82.6 keV for a Thomson scattering x-ray source with a root mean square bandwidth of 1%. Using contrast agents containing gold solutions with Au concentrations ranging from 0.1 to 4.0 wt. %, the performance of the designed imaging system was evaluated by Monte Carlo simulations. Compared to the conventional single-pinhole collimator, the three-pinhole design provides an approximately 2.5-fold enhancement in sensitivity and lowers the detection limit for gold solution from 0.183 to 0.161 wt. %. By increasing the incident x-ray energy to 100 keV, the scatter background was significantly reduced; however, lower contrast-to-noise ratios were observed across all gold solution concentrations, owing to the reduction of fluorescence yield at elevated x-ray energies. For the designed imaging system, a spatial resolution of approximately 0.7 mm can be achieved. Using this system, a complete XFCT scan can be completed in approximately 24 s using a Thomson scattering x-ray source with a photon yield of 1013 photons/s, without inducing pulse pileup effects; meanwhile, gold solutions with a concentration as low as 0.2 wt. % can be clearly resolved at a radiation dose of 2 Gy.Keywords
Funding Information
- National Natural Science Foundation of China (12375157)
- National Natural Science Foundation of China (11905011)
This publication has 41 references indexed in Scilit:
- Simultaneous fluorescence and Compton scattering computed tomography based on linear polarization X-rayNuclear Science and Techniques, 2024
- X-ray polarization characteristics in the nonlinear Thomson scattering of a laser with relativistic electronsNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2022
- Biocompatibility of gold nanoparticles: In-vitro and In-vivo studyMaterials Today: Proceedings, 2021
- Polarized x‐ray excitation for scatter reduction in x‐ray fluorescence computed tomographyMedical Physics, 2018
- Multi-pinhole fluorescent x-ray computed tomography for molecular imagingScientific Reports, 2017
- Dual-energy fluorescent x-ray computed tomography system with a pinhole design: Use of K-edge discontinuity for scatter correctionScientific Reports, 2017
- Highly sensitive detection of the soft tissues based on refraction contrast by in-plane diffraction-enhanced imaging CTNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2008
- Geant4—a simulation toolkitNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2003
- Fluorescence tomography using synchrotron radiation at the NSLSHyperfine Interactions, 1987
- Limits for qualitative detection and quantitative determination. Application to radiochemistryAnalytical Chemistry, 1968