Image quality improvement and motion degradation reduction in shoulder MR imaging: comparison of BLADE and rectilinear techniques at 3-Tesla scanning
Authors:
- 25 June 2022
- journal article
- research article
- Published by Springer Nature in Skeletal Radiology
- Vol. 51 (12) , 2291-2297
Abstract
Objective MR imaging of joints, particularly shoulder, requires a high degree of spatial resolution to ascertain anatomy and pathology. Unfortunately, motion artifacts can reduce the clinical quality of the examinations. BLADE sequence reduces motion degradation improving overall diagnostic imaging quality. The objective was to compare standard, rectilinear k-space coverage turbo spin echo (TSE), and BLADE sequences. Material and methods Over a 4-month period, fifty-seven consecutive patients (22 males, 35 females; mean age: 48.5 years, range: 23–64 years) were scanned using traditional intermediate-weighted spin echo and BLADE sequences. Qualitative evaluation was performed by three musculoskeletal fellowship trained radiologists, each with more than 5 years of experience. Image sequences were evaluated using a Likert scale for each of the following five categories: motion degradation, ghosting/phase misregistration artifacts, star/radial encoding artifacts, fat suppression quality, and overall diagnostic quality. Additionally, image sequences were evaluated for signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) using manually drawn regions of interest (ROI) analysis. Results Ghosting and phase artifacts were lower within BLADE sequence while streak artifacts were higher (p < 0.001). Image fat suppression, tendon and labral appearances, and the overall SNR and CNR were comparable on both sequences (p > 0.05). Conclusion Addition of BLADE reduces motion degradation and improves overall diagnostic imaging quality. Application of BLADE in patient scans suspected of motion artifacts can reduce the frequency of repeat imaging in patients with claustrophobia or those where motion is a concern. By reducing overall imaging time and call backs, it could reduce the cost burden to patients and healthcare system.Keywords
This publication has 12 references indexed in Scilit:
- Efficacy of periodically rotated overlapping parallel lines with enhanced reconstruction (PROPELLER) for shoulder magnetic resonance (MR) imagingEuropean Journal of Radiology, 2016
- Elimination of motion and pulsation artifacts using BLADE sequences in shoulder MR imagingSkeletal Radiology, 2015
- Pairwise Comparison Versus Likert Scale for Biomedical Image AssessmentAmerican Journal of Roentgenology, 2015
- Elimination of motion and pulsation artifacts using BLADE sequences in knee MR imagingMagnetic Resonance Imaging, 2012
- Comparison of Brain MR Images at 1.5T Using BLADE and Rectilinear Techniques for Patients Who Move during Data AcquisitionAmerican Journal of Neuroradiology, 2011
- Comparison of Sagittal T2-Weighted BLADE and Fast Spin-Echo MRI of the Female Pelvis for Motion Artifact and Lesion DetectionAmerican Journal of Roentgenology, 2011
- MRI of the neck at 3 Tesla using the periodically rotated overlapping parallel lines with enhanced reconstruction (PROPELLER) (BLADE) sequence compared with T2‐weighted fast spin‐echo sequenceJournal of Magnetic Resonance Imaging, 2010
- Evaluation of motion correction effect and image quality with the periodically rotated overlapping parallel lines with enhanced reconstruction (PROPELLER) (BLADE) and parallel imaging acquisition technique in the upper abdomenJournal of Magnetic Resonance Imaging, 2008
- Diagnosis and relation to general health of shoulder disorders presenting to primary careRheumatology, 2005
- Motion correction with PROPELLER MRI: Application to head motion and free-breathing cardiac imagingMagnetic Resonance in Medicine, 1999