Accuracy of cardiac output measurements during off-pump coronary artery bypass grafting: according to the vessel anastomosis sites
Authors:
Open Access
- 1 January 2012
- journal article
- Published by The Korean Society of Anesthesiologists in Korean Journal of Anesthesiology
- Vol. 62 (5) , 423-428
Abstract
During beating heart surgery, the accuracy of cardiac output (CO) measurement techniques may be influenced by several factors. This study was conducted to analyze the clinical agreement among stat CO mode (SCO), continuous CO mode (CCO), arterial pressure waveform-based CO estimation (APCO), and transesophageal Doppler ultrasound technique (UCCO) according to the vessel anastomosis sites. This study was prospectively performed in 25 patients who would be undergoing elective OPCAB. Hemodynamic variables were recorded at the following time points: during left anterior descending (LAD) anastomosis at 1 min and 5 min; during obtuse marginal (OM) anastomosis at 1 min and 5 min: and during right coronary artery (RCA) anastomosis at 1 min and 5 min. The variables measured including the SCO, CCO, APCO, and UCCO. CO measurement techniques showed different correlations according to vessel anastomosis site. However, the percent error observed was higher than the value of 30% postulated by the criteria of Critchley and Critchley during all study periods for all CO measurement techniques. In the beating heart procedure, SCO, CCO and APCO showed different correlations according to the vessel anastomosis sites and did not agree with UCCO. CO values from the various measurement techniques should be interpreted with caution during OPCAB.Keywords
This publication has 33 references indexed in Scilit:
- Perioperative Hemodynamic Monitoring with Transesophageal Doppler TechnologyAnesthesia & Analgesia, 2009
- Cardiac Output Measurement in Patients Undergoing Liver Transplantation: Pulmonary Artery Catheter Versus Uncalibrated Arterial Pressure Waveform AnalysisAnesthesia & Analgesia, 2008
- Comparison of Continuous, Stat, and Intermittent Bolus Cardiac Output Measurements during Cardiac SurgeryKorean Journal of Anesthesiology, 2008
- CorrespondenceEuropean Journal of Cardio-Thoracic Surgery, 1998
- The effectiveness of right heart catheterization in the initial care of critically ill patients. SUPPORT InvestigatorsJAMA, 1996
- Semi-continuous versus injectate cardiac output measurement in intensive care patients after cardiac surgeryIntensive Care Medicine, 1996
- Is continuous cardiac output measurement using thermodilution reliable in the critically ill patient?Critical Care Medicine, 1994
- Is continuous cardiac output measurement using thermodilution reliable in the critically ill patient?Critical Care Medicine, 1994
- Hemodynamic status susceptible to slowing of heart rate during thermodilution cardiac output determination in anesthetized patientsCritical Care Medicine, 1990
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986