Evaluation of Implanted Stent Occlusion Status Based on Neointimal Tissue Bioimpedance Simulations
Open Access
- 5 December 2019
- journal article
- research article
- Published by Wiley in Journal of Sensors
- Vol. 2019, 1-11
Abstract
This paper describes the characterization of the light hole, also known as the lumen, in implanted stents affected by restenosis processes using bioimpedance (BI) as a biomarker. The presented approach will enable real-time monitoring of lumens in implanted stents. The basis of the work hereby reported is the fact that neointimal tissues involved in restenosis can be detected and measured through their impedance properties, namely, conductivity and permittivity. To exploit these properties, a 4-electrode setup for BI measurement is proposed. This setup allows study of the influence of the various tissues involved in restenosis fat, muscle, fibre, and endothelium, together with the blood, on the BI value at several frequencies. In addition, BI simulation tests were performed using the electric physics module available in COMSOL Multiphysics®. Interestingly, fat constitutes the most influential layer on the value of impedance (measured in kΩ/μm—magnitude change per micrometre of lumen occlusion). A case study using a standard stent is also presented. In this study, where the involved tissues and blood were simultaneously considered, we conducted an analysis for stable and vulnerable plaques in restenosis test situations. In this regard, the proposed method is useful to test the stent obstruction and detect potential dangerous cases due to nonstable fat accumulation.Keywords
Funding Information
- European Regional Development Fund (RTI2018-093512-B-C21)
- Ministerio de Ciencia, Innovación y Universidades (RTI2018-093512-B-C21)
This publication has 20 references indexed in Scilit:
- The Bio-Oscillator: A Circuit for Cell-Culture AssaysIEEE Transactions on Circuits and Systems II: Express Briefs, 2014
- Identification of Hemodynamically Optimal Coronary Stent Designs Based on Vessel CaliberIEEE Transactions on Biomedical Engineering, 2012
- In-Stent NeoatherosclerosisJACC, 2012
- A novel impedance-based tomography approach for stenotic plaque detection: A simulation studyInternational Journal of Cardiology, 2010
- Towards a self-reporting coronary artery stent—Measuring neointimal growth associated with in-stent restenosis using electrical impedance techniquesBiosensors and Bioelectronics, 2010
- Fully Wireless Implantable Cardiovascular Pressure Monitor Integrated with a Medical StentIEEE Transactions on Biomedical Engineering, 2010
- Comparison of a Polymer-Based Paclitaxel-Eluting Stent With a Bare Metal Stent in Patients With Complex Coronary Artery DiseaseJAMA, 2005
- In vivo intravascular electric impedance spectroscopy using a new catheter with integrated microelectrodesBasic Research in Cardiology, 2004
- Lumen diameter of normal human coronary arteries. Influence of age, sex, anatomic variation, and left ventricular hypertrophy or dilation.Circulation, 1992
- Four-Electrode Null Techniques for Impedance Measurement with High ResolutionReview of Scientific Instruments, 1968