Repeat patient testing‐quality control compared to commercial quality control material for the Sysmex XT‐2000iV hematology analyzer in a multi‐site veterinary laboratory
Open Access
- 14 February 2024
- journal article
- research article
- Published by Wiley in Veterinary Clinical Pathology
- Vol. 53 (S1) , 48-59
Abstract
Background: Quality control material (QCM) for hematology in veterinary laboratories is limited, and repeat patient testing quality control (RPT-QC) is an alternative method using excess matrix-specific samples. Objectives: This study aimed to determine if median differences between RPT-QC analyses for each time interval for RBC, HGB, HCT, and WBC were the same, determine if unified RPT-QC limits can be applied to a network of veterinary laboratories, compare the performance of RPT-QC to commercial QCM for the reference analyzer and evaluate the experience over a 4 month period and design, improve and implement an automated spreadsheet for RPT-QC data management. Methods: The potential to unify individual analyzer RPT-QC limits for red blood cells (RBC), hematocrit (HCT), hemoglobin (HGB), and white blood cells (WBC) on multi-site Sysmex XT-2000-iV analyzers was explored by a difference of means test and confidence interval determination for the median difference for each network analyzer in comparison to the network reference analyzer. User experience of an automated RPT-QC data management Excel spreadsheet was collected by user feedback during monthly meetings. Numbers of out-of-control results and the root causes for these for RPT-QC were compared against those of a commercial QCM over a 4-month period. Results: Differences between individual analyzer RPT-QC limits were too large to allow for unification of network limits. The automated spreadsheet successfully highlighted out-of-control events for RPT-QC. Trends or shifts were more frequent for commercial QCM based on observed performance and a 1-2.5 s QC rule than for RPT-QC. Following routine troubleshooting, RPT-QC out-of-control events were resolved with an alternative RPT-QC sample indicating random error associated with excessive deterioration. Use of an automated spreadsheet for recording RPT-QC, documentation and troubleshooting of out-of-control events, and collating monthly summary calculations were considered an asset in laboratory quality management. Conclusions: RPT-QC can be successfully implemented and integrated into a multi-site veterinary laboratory. Individual analyzer RPT-QC limit generation is recommended. The deterioration of commercial QCM caused shifts or trends in QC results, which initiated more repeat analyses and investigations than did RPT-QC.Keywords
This publication has 14 references indexed in Scilit:
- Evaluation of Sysmex XT‐2000iV analyzer performance across a network of five veterinary laboratories using a commercially available quality control materialVeterinary Clinical Pathology, 2021
- Comparison of traditional statistical quality control using commercially available control materials and two patient‐based quality control procedures for the ADVIA 120 Hematology SystemVeterinary Clinical Pathology, 2018
- Repeat patient testing shows promise as a quality control method for veterinary hematology testingVeterinary Clinical Pathology, 2018
- ASVCP guidelines: Allowable total error hematologyVeterinary Clinical Pathology, 2018
- Comparison of 3 options for choosing control limits in biochemistry testingVeterinary Clinical Pathology, 2017
- Quality Management Recommendations for Automated and Manual In-House Hematology of Domestic AnimalsVeterinary Clinics of North America: Small Animal Practice, 2012
- Toward standardization of quality assessment in laboratory medicine by using the same matrix samples for both internal and external quality assessmentsAccreditation and Quality Assurance, 2010
- Development and Description of the “Random Duplicates” Method of Quality Control for a Hematology LaboratoryAmerican Journal of Clinical Pathology, 1977
- Hematology Control Using Computerized Range ChartLaboratory Medicine, 1972
- Monitoring the Quality of Blood Cell Counts with Replicate Determinations on Routine SamplesAmerican Journal of Clinical Pathology, 1966