  • Comparative Testing of In‐Clinic, Point‐of‐Care Hematology Analyzers With a Commercial Laboratory Reference Analyzer

    by S. Daly, K. P. Freeman, P. A. Graham

    Background: Previous studies have shown that veterinary in‐clinic hematology analyzers may have analytical errors, but few studies have been found that attempt bias correction and look at the effect of this on the achievement of several quality goals and evaluation of analytical performance specifications.Objectives: To design a comparative testing program for use by in‐clinic laboratories with a variety of point‐of‐care (POC) analyzers and to evaluate the performance of in‐clinic hematology analyzers compared to a reference analyzer. To determine whether the removal of bias from the in‐clinic result can be used in preparation for the harmonization of the results with those of the reference laboratory. To determine the analytical performance specifications/quality goals that can be achieved with the removal of bias from the in‐clinic results. To determine whether clinic‐based clinicians and technicians found the information provided valuable for ongoing in‐clinic hematology quality management.Methods: Fresh EDTA whole blood canine samples with results that were within reference limits were processed five times by in‐clinic staff using their in‐clinic analyzer before sending the sample to the reference laboratory for comparison. Results were assessed using quality goals, including total allowable error (TEa) or expert opinion and sigma metrics. The quality goal index (QGI) was calculated for measurands whose performance was < 4 sigma. The potential impact of any difference in results on clinical interpretation was also reviewed.Results: In comparison to the reference analyzer, in‐clinic hematology analyzers often had analytical error and failed to meet quality goals for TEa and sigma metrics. This error was often undetected by the in‐clinic laboratory. Sigma metric < 4 was a useful performance indicator of poor performance, and only 54% of measurands had sigma metrics > 4. Of these poor‐performing instances, 90% were attributed to bias. Performance improved for in‐clinic analyzers by adjusting results to correct for bias in preparation for harmonization.Conclusion: A comparative testing program was successfully designed and implemented to evaluate performance and harmonize in‐clinic hematology analyzers to a reference analyzer using samples that are within reference intervals.


  • Open Access

    Repeat patient testing‐quality control compared to commercial quality control material for the Sysmex XT‐2000iV hematology analyzer in a multi‐site veterinary laboratory

    by S. Daly, M. Rishniw, P. A. Graham, K. P. Freeman

    Background: Quality control material (QCM) for hematology in veterinary laboratories is limited, and repeat patient testing quality control (RPT-QC) is an alternative method using excess matrix-specific samples. Objectives: This study aimed to determine if median differences between RPT-QC analyses for each time interval for RBC, HGB, HCT, and WBC were the same, determine if unified RPT-QC limits can be applied to a network of veterinary laboratories, compare the performance of RPT-QC to commercial QCM for the reference analyzer and evaluate the experience over a 4 month period and design, improve and implement an automated spreadsheet for RPT-QC data management. Methods: The potential to unify individual analyzer RPT-QC limits for red blood cells (RBC), hematocrit (HCT), hemoglobin (HGB), and white blood cells (WBC) on multi-site Sysmex XT-2000-iV analyzers was explored by a difference of means test and confidence interval determination for the median difference for each network analyzer in comparison to the network reference analyzer. User experience of an automated RPT-QC data management Excel spreadsheet was collected by user feedback during monthly meetings. Numbers of out-of-control results and the root causes for these for RPT-QC were compared against those of a commercial QCM over a 4-month period. Results: Differences between individual analyzer RPT-QC limits were too large to allow for unification of network limits. The automated spreadsheet successfully highlighted out-of-control events for RPT-QC. Trends or shifts were more frequent for commercial QCM based on observed performance and a 1-2.5 s QC rule than for RPT-QC. Following routine troubleshooting, RPT-QC out-of-control events were resolved with an alternative RPT-QC sample indicating random error associated with excessive deterioration. Use of an automated spreadsheet for recording RPT-QC, documentation and troubleshooting of out-of-control events, and collating monthly summary calculations were considered an asset in laboratory quality management. Conclusions: RPT-QC can be successfully implemented and integrated into a multi-site veterinary laboratory. Individual analyzer RPT-QC limit generation is recommended. The deterioration of commercial QCM caused shifts or trends in QC results, which initiated more repeat analyses and investigations than did RPT-QC.


  • Repeat patient Testing‐Quality control: An example of implementation in a network of commercial laboratories

    by Kathleen P. Freeman, Susan Daly, Bente Flatland

    The theory and calculations underpinning Repeat Patient Testing-Quality Control (RPT-QC) have been presented in prior publications. This paper gives an example of the process used for implementing RPT-QC in a network of veterinary commercial reference laboratories and the stages associated with the transition to the sole use of RPT-QC. To employ RPT-QC in this commercial laboratory network, eight stages of implementation were identified: (1) education, (2) data collection, (3) calculations, (4) QC recording and documentation, (5) running RPT-QC in parallel with a commercially available quality control material (QCM), (6) development of a Standard Operating Procedure (SOP), (7) development of complementary aspects supporting RPT-QC, and (8) sole use of RPT-QC. Advantages of RPT-QC included cost savings for QCM and External Quality Assessment (EQA) participation and the ability to use commutable specimens with a veterinary matrix at a result level that is of clinical significance for the species. A disadvantage of RPT-QC using a single level of control was the inability to demonstrate stable performance over a range of results. Future avenues for investigation include ongoing refinement of control limits using a pooled standard deviation of the duplicates (SDdup), Sdup over time, investigation of blood samples from species other than the dog, and manipulation of specimens to produce "low abnormal" or "high abnormal" RPT-QC specimens.

  • Open Access

    Repeat patient testing‐quality control with canine samples shows promise as an alternative to commercial quality control material for a network of four Sysmex XT‐2000iV hematology analyzers

    by Susan Daly, Peter A. Graham, Kathleen P. Freeman

    Background: Repeat patient testing‐quality control (RPT‐QC) uses retained patient samples as an alternative to commercial quality control material (QCM). We elected to calculate and validate RPT‐QC limits for red blood cell count (RBC), hemoglobin (HBG), hematocrit (HCT), and white blood cell count (WBC).Objectives: (1) To validate RPT‐QC across a network of four harmonized Sysmex XT‐2000iV hematology analyzers and determine the total error that can be controlled with RPT‐QC. (2) To generate quality control (QC) limits using the standard deviation (SD) of the duplicate measurement differences and determine a suitable simple QC rule with a probability of error detection >0.85 and probability of false rejection 5.5 for RBC, HGB, and WBC, but not for HCT.


  • Open Access

    Quality control validation for a veterinary laboratory network of six Sysmex XT‐2000iV hematology analyzers

    by Susan Daly, Peter A. Graham, Kathleen P. Freeman

    Background: Quality control (QC) validation is an important step in the laboratory harmonization process. This includes the application of statistical QC requirements, procedures, and control rules to identify and maintain ongoing stable analytical performance. This provides confidence in the production of patient results that are suitable for clinical interpretation across a network of veterinary laboratories.Objectives: To determine that a higher probability of error detection (Ped) and lower probability of false rejection (Pfr) using a simple control rule and one level of quality control material (QCM) could be achieved using observed analytical performance than by using the manufacturer's acceptable ranges for QCM on the Sysmex XT‐2000iV hematology analyzers for veterinary use. We also determined whether Westgard Sigma Rules could be sufficient to monitor and maintain a sufficiently high level of analytical performance to support harmonization.Methods: EZRules3 was used to investigate candidate QC rules and determine the Ped and Pfr of manufacturer's acceptable limits and also analyzer‐specific observed analytical performance for each of the six Sysmex analyzers within our laboratory system using the American Society of Veterinary Clinical Pathology (ASVCP)‐recommended or internal expert opinion quality goals (expressed as total allowable error, TEa) as the quality requirement. The internal expert quality goals were generated by consensus of the Quality, Education, Planning, and Implementation (QEPI) group comprised of five clinical pathologists and seven laboratory technicians and managers. Sigma metrics, which are a useful monitoring tool and can be used in conjunction with Westgard Sigma Rules, were also calculated.Results: The QC validation using the manufacturer's acceptable limits for analyzer 1 showed only 3/10 measurands reached acceptable Ped for veterinary laboratories (>0.85). For QC validation based on observed analyzer performance, the Ped was >0.94 using a 1‐2.5s QC rule for the majority of observations (57/60) across the group of analyzers at the recommended TEa. We found little variation in Pfr between manufacturer acceptable limits and individual analyzer observed performance as this is a characteristic of the rule used, not the analyzer performance.Conclusions: An improved probability of error detection and probability of false rejection using a 1‐2.5s QC rule for individual analyzer QC was achieved compared with the use of the manufacturers' acceptable limits for hematology in veterinary laboratories. A validated QC rule (1‐2.5s) in conjunction with sigma metrics (>5.5), desirable bias, and desirable CV based on biologic variation was successful to evaluate stable analytical performance supporting continued harmonization across the network of analyzers.


  • Open Access

    Evaluation of Sysmex XT‐2000iV analyzer performance across a network of five veterinary laboratories using a commercially available quality control material

    by Susan Daly, Kathleen P. Freeman, Peter A. Graham

    Background: Laboratory and instrument harmonization is seldom reported in the veterinary literature despite its advantages to clinical interpretation, including the use of interchangeable results and common reference intervals within a system of laboratories.Objectives: A three‐step process was employed to evaluate and optimize performance and then assess the appropriateness of common reference intervals across a network of six Sysmex XT‐2000iV hematology analyzers at 5 commercial veterinary laboratory sites. The aims were to discover if harmonization was feasible in veterinary hematology and which quality parameters would best identify performance deviations to ensure a harmonized status could be maintained.Methods: The performance of 10 measurands of a commercially available quality control material (Level 2—Normal e‐CHECK (XE)‐Hematology Control) was evaluated during three 1‐month time periods. Precision and bias were assessed with Six Sigma, American Society of Veterinary Clinical Pathology (ASVCP) total error quality goals and biologic variation (BV)‐based quality goal approaches to performance measurement.Results: Instrument adjustments were made to 1 analyzer twice and 3 analyzers once between evaluations to improve performance and achieve harmonization. Sigma metrics improved from 37/50 > 6 to 58/60 > 6 and to all >5 over the course of the harmonization project. BV‐based quality goals for desirable bias and for laboratory systems of 0.33 × CVI (within‐subject biologic variation) were more sensitive and useful for assessing performance than the ASVCP total error goals.Conclusions: Optimization and harmonization were achieved, and because BV‐derived bias goals were achieved, common reference intervals could be implemented across the network of analyzers.