Effects of charge carrier trapping on polycrystalline PbO x-ray imaging detectors
Author:
- 1 October 2008
- journal article
- conference paper
- Published by AIP Publishing in Journal of Applied Physics
- Vol. 104 (7) , 074506
Abstract
The effects of charge carrier trapping on the x-ray sensitivity, resolution, and detective quantum efficiency (DQE) in polycrystalline lead oxide (PbO) x-ray imaging detectors are theoretically analyzed. The theoretical models for calculating carrier trapping-limited x-ray sensitivity and modulation transfer function (MTF), and DQE incorporating polyenergetic x-ray spectrum are described. A cascaded linear system model is developed for calculating the spatial frequency dependent DQE of PbO detectors by incorporating the effects of bulk charge carrier trapping on the MTF and the x-ray interaction depth dependent charge collection efficiency and noise. The theoretical calculations are compared with the published experimental data and show a very good agreement. From the fittings of the sensitivity and MTF curves, the electron and hole ranges in polycrystalline PbO are found to be and , respectively. The dependent noise power spectrum and performances are analyzed as a function of applied electric field and carrier lifetimes. The calculation shows that the carrier trapping is responsible for a 21% reduction in the MTF at the applied field of at the Nyquist frequency and the aliased NPS at is reduced to 63% of that at zero spatial frequency. The sensitivity, MTF, , and strongly depend on the applied electric field and carrier ranges. It is found that the for fluoroscopic applications can be improved dramatically by simply enhancing the carrier lifetimes through improving material properties of the PbO sample.
Keywords
This publication has 16 references indexed in Scilit:
- The effects of large signals on charge collection in radiation detectors: Application to amorphous selenium detectorsJournal of Applied Physics, 2006
- Analysis of lead oxide (PbO) layers for direct conversion X-ray detectionIEEE Transactions on Nuclear Science, 2005
- Evaluation of the imaging properties of an amorphous selenium‐based flat panel detector for digital fluoroscopyMedical Physics, 2004
- Modulation transfer function of photoconductive x-ray image detectors: effects of charge carrier trappingJournal of Physics D: Applied Physics, 2003
- DQE of photoconductive x-ray image detectors: application to a-SeJournal of Physics D: Applied Physics, 2002
- Direct conversion detectors: The effect of incomplete charge collection on detective quantum efficiencyMedical Physics, 2002
- Charge collection and absorption-limited sensitivity of x-ray photoconductors: Applications to a-Se and HgI2Applied Physics Letters, 2002
- X-ray sensitivity of photoconductors: application to stabilized a-SeJournal of Physics D: Applied Physics, 2000
- Review X-ray photoconductors and stabilized a-Se for direct conversion digital flat-panel X-ray image-detectorsJournal of Materials Science: Materials in Electronics, 2000
- X‐ray imaging using amorphous selenium: Inherent spatial resolutionMedical Physics, 1995