Reliability of uptake estimates in FDG PET as a function of acquisition and processing protocols using the CPET
Authors:
- M. SoretMSM. Soret
- O. de DreuilleODO. de Dreuille
- H. FoehrenbachHFH. Foehrenbach
- I. BuvatIBI. Buvat
- 5 December 2005
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Nuclear Science
- Vol. 52 (5) , 1447-1452
Abstract
Standardized uptake values (SUV) are commonly used in FDG PET to characterize suspicious high uptakes. To better understand the reliability and the limits of SUV, we studied the accuracy of uptake estimates as a function of a number of parameters using phantom data acquired on the CPET. Methods: Using the Data Spectrum thorax phantom in which spheres were inserted, we studied the effect of the sphere sizes, of out-of-the-field-of-view activity (OFOVA), of the emission scan duration and of attenuation and partial volume effect (PVE) corrections upon biases in tumor-to-normal tissue uptake ratio (TNR) estimates. Considering a specific acquisition and processing protocol as a reference, we determined the changes in TNR when modifying this protocol. Results: For a true TNR of 8, estimated TNR were strongly dependent on lesion size, but also on the acquisition and processing protocol. Depending on the method used to derive the attenuation map, estimated TNR could change by more than 50% for lung spheres less than 2 cm in diameter. Using PVE correction, TNR increased by a factor greater than 2 for spheres less than 2 cm in diameter. The very method used to estimate sphere activity from the reconstructed images could also change the TNR by more than 50%. Even with CT-based attenuation correction and PVE correction, TNR estimates remained underestimated by more than 20% in lesions less than 2 cm in diameter. Conclusion: Biases in uptake estimates strongly depend on the acquisition and processing protocols. This suggests that comparing SUV between studies make sense only if the scanner and the acquisition and processing protocols are strictly identical.Keywords
This publication has 11 references indexed in Scilit:
- Unified description and validation of Monte Carlo simulators in PETPhysics in Medicine & Biology, 2005
- Positron Emission Tomography in Non–Small-Cell Lung Cancer: Prediction of Response to Chemotherapy by Quantitative Assessment of Glucose UseJournal of Clinical Oncology, 2003
- Prognostic value of [ 18 F]FDG-PET imaging in small cell lung cancerEuropean Journal of Nuclear Medicine and Molecular Imaging, 2003
- CT-based attenuation correction in the calculation of semi-quantitative indices of [18F]FDG uptake in PETEuropean Journal of Nuclear Medicine and Molecular Imaging, 2002
- Biases affecting the measurements of tumor-to-background activity ratio in PETIEEE Transactions on Nuclear Science, 2002
- Influence of OSEM and segmented attenuation correction in the calculation of standardised uptake values for [18F]FDG PETEuropean Journal of Nuclear Medicine and Molecular Imaging, 2001
- CT assessment of tumour response to treatment: comparison of linear, cross-sectional and volumetric measures of tumour size.British Journal of Radiology, 2000
- Exact and approximate rebinning algorithms for 3-D PET dataIEEE Transactions on Medical Imaging, 1997
- Analysis of Emission Tomographic Scan Data: Limitations Imposed by Resolution and BackgroundJournal of Computer Assisted Tomography, 1984