Comparison of Flow Characteristics of Novel Three-Dimensional Printed Ureteral Stents Versus Standard Ureteral Stents in a Porcine Model

Authors:

Abstract
Purpose: We compared the flow characteristics of novel three-dimensional (3D) printed ureteral stents with four conventional double-pigtail stents in an ex vivo porcine model. Materials and Methods: In six ex vivo porcine urinary systems with kidneys and ureters intact, we deployed a 5F occlusion catheter in an interpolar calix. We tested each system with antegrade irrigation with a 0.9% saline bag placed 35 cm above the renal pelvis. We evaluated four standard stents (6F Universa® Soft, 7F Percuflex,™ 7/10F Applied Endopyelotomy, 8.5F Filiform Double Pigtail) and compared them with a 9F 3D printed prototype stent. For each stent, we measured the total, extraluminal, and intraluminal flow rates. Results: The mean total flow rates for 3D printed stents were significantly higher than the 6F, 7F, and 7/10F stents (PPPPPConclusions: In this pilot study, 3D printed stents manifested a mean total flow rate comparable to the flow rates of contemporary stents. Continued advances in technology and material may permit functionally feasible 3D printed ureteral stents.