Accuracy of Optical Heart Rate Measurements for 10 Commercial Wearables in Different Climate Conditions and Activities: Instrument Validation Study
Top Cited Papers
Open Access
- 17 February 2026
- journal article
- research article
- Published by JMIR Publications Inc. in JMIR Formative Research
- Vol. 10, e85186
Abstract
Background: Commercial wearable devices allow for continuous heart rate (HR) monitoring in daily life. Their accuracy under ecologically valid conditions, however, remains insufficiently independently tested, especially during irregular activity, cognitive stress, and variable climates. Objective: This study evaluated the HR accuracy of 10 commercially available wearables under controlled variations in physical activity, cognitive stress, and temperature. We hypothesized that physical activity irregularity, cognitive stress, and thermal climate conditions would affect measurement accuracy. Methods: Forty-five healthy adults (21‐68, mean 34, SD 12 y) completed a standardized protocol in climate-controlled chambers simulating neutral (23 °C), hot (36 °C), and cold (10 °C) conditions. Tasks included rest, cognitive stress (Montreal Imaging Stress Task), steady walking, and intermittent walking. Each of the 10 devices (Fitbit Charge 6, Fitbit Inspire 3, Garmin Vivosmart 5, Garmin Vivoactive 5, Apple Watch SE, Google Pixel Watch 2, Polar Ignite 3, Polar Pacer, Xiaomi Watch 2, and Oura Ring Gen 3) was compared against electrocardiogram-derived HR from a Zephyr BioHarness chest strap. Accuracy was assessed using mean absolute error (MAE), mean absolute percentage error (MAPE), repeated-measures concordance correlation coefficient (CCC), and Bland-Altman analysis. Results: Significant variability across the devices was observed. Fitbit Charge 6 (MAE 4.5 bpm, MAPE 5.5%, CCC 0.93) and Google Pixel Watch 2 (MAE 4.9 bpm, MAPE 6.7%, CCC 0.87) showed strong agreement with the gold standard. In contrast, Fitbit Inspire 3, Polar Ignite 3, Polar Pacer, and Oura Ring displayed larger errors (MAE 9‐14 bpm, MAPE 11%‐16%) and lower CCC values (0.45‐0.66). The climate conditions did not significantly affect the measurement accuracy of the test devices. The activity type, however, did have a significant effect: intermittent walking increased errors for multiple devices. Conclusions: Wearable HR measurement accuracy is device-specific and context-dependent. Moderate climates did not impair performance, but irregular movement reduced accuracy. Fitbit Charge 6 and Google Pixel Watch 2 demonstrated the highest reliability, supporting their use in health and sports monitoring. Careful device selection and context-aware interpretation remain critical for applied and clinical applications.Keywords
This publication has 38 references indexed in Scilit:
- GAIDeT (Generative AI Delegation Taxonomy): A taxonomy for humans to delegate tasks to generative artificial intelligence in scientific research and publishingAccountability in Research, 2025
- Optimisation and pre-use suitability selection for wrist photo-plethysmography based heart rate monitoring in cardiac patientsEuropean Heart Journal - Digital Health, 2025
- Systematic review and meta-analysis of performance of wearable artificial intelligence in detecting and predicting depressionnpj Digital Medicine, 2023
- Investigating sources of inaccuracy in wearable optical heart rate sensorsnpj Digital Medicine, 2020
- Accuracy of Wrist-Worn Heart Rate MonitorsJAMA Cardiology, 2017
- The Concordance Correlation Coefficient for Repeated Measures Estimated by Variance ComponentsJournal of Biopharmaceutical Statistics, 2009
- Agreement Between Methods of Measurement with Multiple Observations Per IndividualJournal of Biopharmaceutical Statistics, 2007
- The effect of local mild cold exposure on pulse transit timePhysiological Measurement, 2006
- Vagal influence on working memory and attentionInternational Journal of Psychophysiology, 2003
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986