Optimisation and pre-use suitability selection for wrist photo-plethysmography based heart rate monitoring in cardiac patients
Open Access
- 23 July 2025
- journal article
- Published by Oxford University Press (OUP) in European Heart Journal - Digital Health
- Vol. 6 (5) , 1024-1035
Abstract
Introduction: Sensor placement, activity type influencing wrist movements, and individual characteristics impact accuracy of wrist-worn photoplethysmography (PPG)-based heart rate (HR) monitors. This study investigated technical interventions to optimise PPG accuracy in patients with cardiac disease. Methods: The Fitbit Inspire 2 PPG monitor was evaluated across three cohorts, using a Polar H10 chest strap as reference: (1) 10 healthy volunteers performed wrist movements with the monitor placed one or three fingers above the wrist to identify optimal placement; (2) 10 volunteers engaged in sport activities (walking, running, cycling, rowing); (3) 30 cardiac rehabilitation patients were monitored during exercise to assess baseline accuracy. Patients with low accuracy (mean absolute percentage error (MAPE) <10% for <70% of training time) underwent technical interventions (sensor cleaning, forearm shaving, position fixation, and/or relocation to the volar wrist). Results: Placement three vs. one fingers above the wrist was significantly more accurate (mean difference in MAPE: -11.4%, p<0.001). Walking showed the highest accuracy (MAPE=3.8%), followed by cycling (MAPE=6.9%) and running (MAPE=8.5%), while rowing had the lowest accuracy (MAPE=13.4%, p<0.001). Among CR patients, 66.7% achieved high baseline accuracy. Technical interventions improved accuracy in 50.0% of those with low baseline accuracy, but no significant predictors of optimisation success were identified. Conclusions: Accurate PPG-based monitoring requires a sensor placed higher on the wrist. Nevertheless, only two-thirds of patients are suitable for such monitoring, with improvement by technical adaptations possible (but impractical) in the others. Therefore, assessing baseline accuracy is a prerequisite before relying on these devices for activity guidance.Keywords
This publication has 26 references indexed in Scilit:
- Accuracy of Consumer Wearable Heart Rate Measurement During an Ecologically Valid 24-Hour Period: Intraindividual Validation StudyJMIR mHealth and uHealth, 2019
- Methodological Shortcomings of Wrist-Worn Heart Rate Monitors ValidationsJournal of Medical Internet Research, 2018
- Validity of Wearable Activity Monitors during Cycling and Resistance ExerciseMedicine & Science in Sports & Exercise, 2018
- Comparison of Polar M600 Optical Heart Rate and ECG Heart Rate during ExerciseMedicine & Science in Sports & Exercise, 2017
- Variable Accuracy of Wearable Heart Rate Monitors during Aerobic ExerciseMedicine & Science in Sports & Exercise, 2017
- Forearm hair density and risk of keratinocyte cancers in Australian adultsArchives of Dermatological Research, 2016
- Characterization and reduction of motion artifacts in photoplethysmographic signals from a wrist-worn device2015
- Validation of photoplethysmography as a method to detect heart rate during rest and exerciseJournal of Medical Engineering & Technology, 2015
- Relationship Between Measurement Site and Motion Artifacts in Wearable Reflected PhotoplethysmographyJournal of Medical Systems, 2010
- The validity and practicality of sun-reactive skin types I through VIArchives of Dermatology, 1988