Comparison of quantitative regional perfusion‐weighted phase resolved functional lung (PREFUL) MRI with dynamic gadolinium‐enhanced regional pulmonary perfusion MRI in COPD patients
- 22 October 2018
- journal article
- research article
- Published by Wiley in Journal of Magnetic Resonance Imaging
- Vol. 49 (4) , 1122-1132
Abstract
Background: Perfusion‐weighted noncontrast‐enhanced proton lung MRI during free breathing is maturing as a novel technique for assessment of regional lung perfusion, but has not yet been validated in chronic obstructive pulmonary disease (COPD) patients.Purpose: To compare pulmonary parenchymal perfusion assessed by noncontrast‐enhanced perfusion‐weighted phase‐resolved functional lung (PREFUL)‐MRI with lung perfusion determined with dynamic gadolinium‐enhanced (DCE)‐MRI and with lung function test parameters.Study Type: Prospective.Population: A single‐center subset of the COPD cohort "COPD and SYstemic consequenzes‐COmorbidities NETwork" (COSYCONET). Forty‐seven patients with COPD (median age 66 [57–70] years) were studied.Field Strength/Sequence: For PREFUL‐MRI a spoiled gradient echo sequence and for DCE‐MRI, a 3D time‐resolved spoiled gradient echo sequence was used at 1.5T.Assessment: PREFUL‐MRI coronal slices were acquired in free breathing. DCE‐MRI was performed in breath‐hold with administration of 0.025 mmol/kg bodyweight of gadobutrol i.v. at a rate of 4 ml/s and pulmonary blood flow (PBF) maps were calculated. Slices of PREFUL and DCE‐MRI were matched by their ventrodorsal position and corresponding slices were coregistered for evaluation. Perfusion defect percentages (QDP) were calculated for both methods.Statistical Tests: The obtained parameters were correlated using Spearman's correlation coefficient (r) and Bland–Altman plot analysis.Results: PREFUL‐QDP showed an absolute and spatial agreement with PBF‐QDP on a global (39.3 (31.8–45.5)% vs. 44.7 (35.4–50.0)% with a spatial overlap of 62.2 (57.2–67.2)%)) as well as on a lobar level and correlated with lung function test parameters (PREFUL‐QDP vs. FEV1, r = –0.75, P < 0.0001). There was a systematic overestimation of PREFUL‐QDP compared with PBF‐QDP, mainly in the lower lobes, resulting in an overall overestimation for the whole lung with a mean difference of 5% (95% confidence interval [CI]: 3.0%; 7.0%; STD 6.8%).Data Conclusion: PREFUL‐MRI is a promising noninvasive, radiation‐free tool for quantification of regional perfusion in COPD patients.Level of Evidence: 1Technical Efficacy: Stage 2J. Magn. Reson. Imaging 2019;49:1122–1132.Keywords
This publication has 35 references indexed in Scilit:
- Quantitative and Semiquantitative Measures of Regional Pulmonary Microvascular Perfusion by Magnetic Resonance Imaging and Their Relationships to Global Lung Perfusion and Lung Diffusing CapacityInvestigative Radiology, 2013
- Computed tomography–based biomarker provides unique signature for diagnosis of COPD phenotypes and disease progressionNature Medicine, 2012
- Quantitative Computed Tomography Measures of Emphysema and Airway Wall Thickness Are Related to Respiratory SymptomsAmerican Journal of Respiratory and Critical Care Medicine, 2010
- Non‐contrast‐enhanced perfusion and ventilation assessment of the human lung by means of fourier decomposition in proton MRIMagnetic Resonance in Medicine, 2009
- Quantitative contrast‐enhanced perfusion measurements of the human lung using the prebolus approachJournal of Magnetic Resonance Imaging, 2009
- Pixel-by-pixel deconvolution of bolus-tracking data: optimization and implementationPhysics in Medicine & Biology, 2006
- Projections of Global Mortality and Burden of Disease from 2002 to 2030PLOS Medicine, 2006
- Computed Tomographic Measurements of Airway Dimensions and Emphysema in SmokersAmerican Journal of Respiratory and Critical Care Medicine, 2000
- Morphometric analysis of white matter lesions in MR images: method and validationIEEE Transactions on Medical Imaging, 1994
- Measures of the Amount of Ecologic Association Between SpeciesEcology, 1945