Fixation Failures of Dual Mobility Cups: A Mid-term Study of 2601 Hip Replacements
Authors:
- 1 July 2012
- journal article
- research article
- Published by Wolters Kluwer Health in Clinical Orthopaedics and Related Research
- Vol. 470 (7) , 1932-1940
Abstract
The use of dual-mobility cups has increased because of a low rate of dislocations combined with a 96% 15-year survival rate. However, late cup migrations have been attributed to their fixation (tripod - exact fit with two pegs and one extraacetabular screw) and the absence of porous coating. In a second-generation device, the designs were modified to achieve press-fit fixation and a layer of titanium beads was sintered on stainless steel cups. We therefore (1) determined the midterm survival of press-fit, grit-blasted, second-generation cups with or without additional screws, compared with original tripod and (2) compared survival of grit-blasted dual-mobility cups with bimetallic porous-coated cups. From a multiinstitutional trial, we reviewed 2408 patients with osteoarthritis implanted with 2601 prostheses of seven designs of a second-generation dual-mobility cup. The criteria for failure were migration, widening radiolucencies in any zone of the interface, or revision for cup loosening. The minimum followup was 5 years (mean, 7.7 years; range, 5–11 years). The 8-year survival rate of press-fit, grit-blasted cups was lower than that for press-fit, grit-blasted cups fixed with screws (91% versus 100%) and for tripod fixation (98%). The 8-year survival rate of press-fit, grit-blasted cups was less than that for press-fit, porous-coated cups made of the same alloy (91% versus 95%). The data suggested primary fixation of grit-blasted dual-mobility cups should be secured with screws. Porous coating sintered on the convex side improved midterm survivorship. No deleterious effect of metallosis resulted from sintered titanium beads on stainless steel. Long-term followup is required to confirm these findings. Level IV, therapeutic study. See Guidelines for Authors for a complete description of levels of evidence.Keywords
This publication has 32 references indexed in Scilit:
- The dual mobility socket concept: experience with 668 casesInternational Orthopaedics, 2010
- Ten-year results with the Morscher press-fit cup: an uncemented, non-modular, porous-coated cup inserted without screwsInternational Orthopaedics, 2010
- Use of a Dual Mobility Socket to Manage Total Hip Arthroplasty InstabilityClinical Orthopaedics and Related Research, 2009
- The use of a dual-articulation acetabular cup system to prevent dislocation after primary total hip arthroplasty: analysis of 384 cases at a mean follow-up of 15 yearsInternational Orthopaedics, 2008
- Grit-blasted and hydroxyapatite-coated total hip arthroplasty: an 11- to 14-year follow-up studyJournal of Orthopaedic Science, 2004
- Classification of Surgical ComplicationsAnnals of Surgery, 2004
- Failure of hydroxyapatite-coated acetabular cupsThe Journal of Bone and Joint Surgery. British volume, 2002
- Failure of HA coating on a gritblasted acetabular cup: 155 patients followed for 7-10 yearsActa Orthopaedica, 2002
- Survival probabilities (the Kaplan-Meier method)BMJ, 1998
- Accuracy of migration analysis in hip arthroplasty Digitized and conventional radiography, compared to radiostereometry in 51 patientsActa Orthopaedica, 1995