Long-term results of high vs. normal impedance ventricular leads on actual (Real-Life) pacemaker generator longevity
Open Access
- 4 November 2008
- journal article
- research article
- Published by Oxford University Press (OUP) in EP Europace
- Vol. 11 (2) , 200-205
Abstract
The long-term effects of high impedance vs. standard impedance pacing leads on actual generator longevity were studied. In 40 patients (21 females, age 73 ± 13 years) with standard dual-chamber pacemaker indication, bipolar standard impedance ventricular leads and high-impedance leads were implanted in a randomized fashion. Identical pacemaker generators and atrial pacing leads were implanted in all patients. Patients were observed during a mean follow-up of 89.8 ± 8.8 months before pacemaker replacement. Initially, the patients who received the high-impedance leads had a lower current drain as compared with standard pacing impedance leads, and the estimated pacemaker longevity was significantly prolonged, too. But this pattern disappeared after 6 years of follow-up, and finally the actual pacemaker generators’ replacement time was 86.7 ± 6.8 months in standard impedance lead group vs. 91.2 ± 10.3 months in high-impedance lead group ( P = 0.17). Implantation of high pacing impedance leads for ventricular stimulation does not result in a benefit with respect to pacemaker longevity as compared with standard impedance leads.Keywords
This publication has 11 references indexed in Scilit:
- Pacemaker Longevity: Are We Getting What We Are Promised?Pacing and Clinical Electrophysiology, 2006
- Role of ventricular autocapture function in increasing longevity of DDDR pacemakers: a prospective studyEP Europace, 2006
- Long‐Term Performance of Transvenous, Steroid‐Eluting, High Impedance, Passive‐Fixation Ventricular Pacing LeadsPacing and Clinical Electrophysiology, 2004
- Prospects for Biological Cardiac Pacemaker SystemsPacing and Clinical Electrophysiology, 2003
- Hochohm-ElektrodenHerzschrittmachertherapie + Elektrophysiologie, 2001
- Extension of Generator Longevity by Use of High Impedance Ventricular LeadsPacing and Clinical Electrophysiology, 2001
- Stable Electrical Performance of High Efficiency Pacing Leads Having Small Surface, Steroid‐Eluting Pacing ElectrodesPacing and Clinical Electrophysiology, 1999
- Steroid Eluting High Impedance Pacing Leads Decrease Short and Long‐Term Current Drain: Results from a Multicenter Clinical Trial.Pacing and Clinical Electrophysiology, 1999
- ACC/AHA Guidelines for Implantation of Cardiac Pacemakers and Antiarrhythmia Devices: Executive SummaryCirculation, 1998
- Initial Experience with 1.5‐mm2 High Impedance, Steroid‐luting Pacing ElectrodesPacing and Clinical Electrophysiology, 1996