Thermal considerations for microswimmer trap-and-release using standing surface acoustic waves
- 7 May 2021
- journal article
- research article
- Published by Royal Society of Chemistry (RSC) in Lab on a Chip
- Vol. 21 (13) , 2534-2543
Abstract
Controlled trapping of cells and microorganisms using substrate acoustic waves (SAWs; conventionally termed surface acoustic waves) has proven useful in numerous biological and biomedical applications owing to the label- and contact-free nature of acoustic confinement. However, excessive heating due to vibration damping and other system losses potentially compromises the biocompatibility of the SAW technique. Herein, we investigate the thermal biocompatibility of polydimethylsiloxane (PDMS)-based SAW and glass-based SAW [that supports a bulk acoustic wave (BAW) in the fluid domain] devices operating at different frequencies and applied voltages. First, we use infrared thermography to produce heat maps of regions of interest (ROI) within the aperture of the SAW transducers for PDMS- and glass-based devices. Motile Chlamydomonas reinhardtii algae cells are then used to test the trapping performance and biocompatibility of these devices. At low input power, the PDMS-based SAW system cannot generate a large enough acoustic trapping force to hold swimming C. reinhardtii cells. At high input power, the temperature of this device rises rapidly, damaging (and possibly killing) the cells. The glass-based SAW/BAW hybrid system, on the other hand, can not only trap swimming C. reinhardtii at low input power, but also exhibits better thermal biocompatibility than the PDMS-based SAW system at high input power. Thus, a glass-based SAW/BAW device creates strong acoustic trapping forces in a biocompatible environment, providing a new solution to safely trap active microswimmers for research involving motile cells and microorganisms.Keywords
Funding Information
- Division of Chemical, Bioengineering, Environmental and Transport Systems (CBET-1944063)
- Division of Civil, Mechanical and Manufacturing Innovation (CMMI-1633971)
- Washington University in St. Louis
This publication has 88 references indexed in Scilit:
- Isolation and retrieval of circulating tumor cells using centrifugal forcesScientific Reports, 2013
- Circulating Tumor Cell Microseparator Based on Lateral Magnetophoresis and Immunomagnetic NanobeadsAnalytical Chemistry, 2013
- Flow-switching allows independently programmable, extremely stable, high-throughput diffusion-based gradientsLab on a Chip, 2013
- On-chip manipulation of single microparticles, cells, and organisms using surface acoustic wavesProceedings of the National Academy of Sciences of the United States of America, 2012
- Automated and temperature-controlled micro-PIV measurements enabling long-term-stable microchannel acoustophoresis characterizationLab on a Chip, 2011
- Imaging Single-Cell Signaling Dynamics with a Deterministic High-Density Single-Cell Trap ArrayAnalytical Chemistry, 2011
- Particle manipulation in a microfluidic channel using acoustic trapBiomedical Microdevices, 2011
- High-Performance Single Cell Genetic Analysis Using Microfluidic Emulsion Generator ArraysAnalytical Chemistry, 2010
- An Equilibrium Method for Continuous-Flow Cell Sorting Using DielectrophoresisAnalytical Chemistry, 2008
- Electrosonic ejector microarray for drug and gene deliveryBiomedical Microdevices, 2007