Advantages and limitations of noninvasive devices for cardiac output monitoring: a literature review
- 7 April 2023
- journal article
- review article
- Published by Wolters Kluwer Health in Current Opinion in Critical Care
- Vol. 29 (3) , 259-267
Abstract
Purpose of review: Individualized hemodynamic optimization often requires cardiac output monitoring, mostly for goal-directed therapy in the operating room and fluid responsiveness assessment in the intensive care unit. Different technologies for noninvasive cardiac output determination have become available over the recent years. It is therefore crucial for care providers to be made aware of the advantages and limitations of those different devices for an appropriate use at the bedside. Recent findings: Nowadays, different noninvasive technologies exist, each with their advantages and limitations, but none are considered interchangeable with bolus thermodilution. Different clinical studies however, emphasize that the trending ability of such devices allows to guide decisions from care providers, and that their use may be associated with improved prognosis, especially in the operating room. Recent studies have also reported their potential use for hemodynamic optimization in specific populations. Summary: Noninvasive cardiac output monitoring may have a clinical impact on patient outcomes. Further studies are required to evaluate their clinical relevance, notably in the intensive care unit. Noninvasive monitoring opens up the possibility for hemodynamic optimization in specific or low-risk populations, the benefit of which remains to be assessed.Keywords
This publication has 65 references indexed in Scilit:
- A Double-Blind Randomized Controlled Clinical Trial to Assess the Effect of Doppler Optimized Intraoperative Fluid Management on Outcome Following Radical CystectomyJournal of Urology, 2011
- The Impact of Phenylephrine, Ephedrine, and Increased Preload on Third-Generation Vigileo-FloTrac and Esophageal Doppler Cardiac Output MeasurementsAnesthesia & Analgesia, 2011
- Assessment of Trending Ability of Cardiac Output Monitors by Polar Plot MethodologyJournal of Cardiothoracic and Vascular Anesthesia, 2011
- Minimally Invasive Measurement of Cardiac Output during Surgery and Critical CareAnesthesiology, 2010
- Bench-to-bedside review: The importance of the precision of the reference technique in method comparison studies – with specific reference to the measurement of cardiac outputCritical Care, 2009
- Noninvasive cardiac output monitoring (NICOM): a clinical validationIntensive Care Medicine, 2007
- Randomized clinical trial assessing the effect of Doppler-optimized fluid management on outcome after elective colorectal resectionBritish Journal of Surgery, 2006
- The validity of trans-esophageal Doppler ultrasonography as a measure of cardiac output in critically ill adultsIntensive Care Medicine, 2004
- Randomised controlled trial assessing the impact of a nurse delivered, flow monitored protocol for optimisation of circulatory status after cardiac surgeryBMJ, 2004
- Thoracic Electrical Bioimpedance Measurement of Cardiac OutputCritical Care Medicine, 1994