The choice of IOP measurement method for assessing the effectiveness of glaucoma surgery
Open Access
- 12 May 2023
- journal article
- research article
- Published by Research Institute of Eye Diseases in National Journal glaucoma
- Vol. 22 (2) , 39-43
Abstract
Purpose of this study — to compare the results of different tonometry methods before surgical treatment of glaucoma and in the early postoperative period.The study was conducted on a group of 50 patients (50 eyes) aged 55 to 80 years with uncompensated primary open-angle glaucoma, who were admitted to in-patient department for glaucoma surgery. Patients were examined using bidirectional applanation tonometry of the cornea performed on Ocular Response Analyzer, pneumotonometry on Canon TX-20P device, and with Icare tonometer. These studies were carried out on the day before the surgery, the next day, and 2 weeks after the operation.Significant differences in tonometry readings were revealed between all tested devices at high intraocular pressure (IOP) levels (before glaucoma surgery). Significant differences were also found in IOP values obtained with Icare tonometer in the central zone of the cornea and in the middle periphery in the nasal and temporal sectors. A significant difference between the indicators remained on the next day after surgery, except for the Icare readings. After two weeks, the tonometric parameters did not differ significantly from each other.Corneal compensated IOP (IOPcc) is the most important tonometric indicator in clinical practice because it takes into account the individual biomechanical characteristics of the patient’s cornea. When examining patients with glaucoma, the IOPcc indicator significantly differed in uncompensated IOP, which is important for determining the correct treatment tactics. When assessing the level of IOP after surgery this trend persisted, indicating a systematic underestimation of IOP level (overestimation of the effect of glaucoma surgery). The reliability of the study is confirmed by the results of measurements on unoperated fellow eyes (control).Keywords
This publication has 4 references indexed in Scilit:
- Influence of prostaglandin analogues on intraocular pressure fluctuations in body position changeRussian Journal of Clinical Ophthalmology, 2022
- Coefficient of biomechanical stress in assessment of the degree of intraocular pressure compensationRussian Annals of Ophthalmology, 2021
- Ophthalmic risk factors for primary open angle glaucomaRussian Journal of Clinical Ophthalmology, 2019
- Accuracy of the ICare rebound tonometer in glaucomatous eyes with topical ocular hypotensive medicationOphthalmic and Physiological Optics, 2008