Evaluation of Non-Uniform Image Quality Caused by Anode Heel Effect in Digital Radiography Using Mutual Information
Open Access
Abstract
Anode heel effects are known to cause non-uniform image quality, but no method has been proposed to evaluate the non-uniform image quality caused by the heel effect. Therefore, the purpose of this study was to evaluate non-uniform image quality in digital radiographs using a novel circular step-wedge (CSW) phantom and normalized mutual information (nMI). All X-ray images were acquired from a digital radiography system equipped with a CsI flat panel detector. A new acrylic CSW phantom was imaged ten times at various kVp and mAs to evaluate overall and non-uniform image quality with nMI metrics. For comparisons, a conventional contrast-detail resolution phantom was imaged ten times at identical exposure parameters to evaluate overall image quality with visible ratio (VR) metrics, and the phantom was placed in different orientations to assess non-uniform image quality. In addition, heel effect correction (HEC) was executed to elucidate the impact of its effect on image quality. The results showed that both nMI and VR metrics significantly changed with kVp and mAs, and had a significant positive correlation. The positive correlation is suggestive that the nMI metrics have a similar performance to the VR metrics in assessing the overall image quality of digital radiographs. The nMI metrics significantly changed with orientations and also significantly increased after HEC in the anode direction. However, the VR metrics did not change significantly with orientations or with HEC. The results indicate that the nMI metrics were more sensitive than the VR metrics with regards to non-uniform image quality caused by the anode heel effect. In conclusion, the proposed nMI metrics with a CSW phantom outperformed the conventional VR metrics in detecting non-uniform image quality caused by the heel effect, and thus are suitable for quantitatively evaluating non-uniform image quality in digital radiographs with and without HEC.Keywords
Funding Information
- Ministry of Science and Technology (MOST107-2314-B-037-050-MY2)
This publication has 25 references indexed in Scilit:
- Effect of anode angle on photon beam spectra and depth dose characteristics for X-RAD320 orthovoltage unitReports of Practical Oncology and Radiotherapy, 2013
- Mutual information and redundancy in spontaneous communication between cortical neuronsBiological Cybernetics, 2011
- Mutual information-based evaluation of image quality with its preliminary application to assessment of medical imaging systemsJournal of Electronic Imaging, 2009
- Detector or System? Extending the Concept of Detective Quantum Efficiency to Characterize the Performance of Digital Radiographic Imaging SystemsRadiology, 2008
- An Automatic Correction Method for the Heel Effect in Digitized Mammography ImagesJournal of Digital Imaging, 2007
- Information Entropy Measure for Evaluation of Image QualityJournal of Digital Imaging, 2007
- Comparison of visual grading analysis and determination of detective quantum efficiency for evaluating system performance in digital chest radiographyEuropean Radiology, 2003
- Comparison of computed radiography and film/screen combination using a contrast-detail phantomJournal of Applied Clinical Medical Physics, 2003
- Investigation of basic imaging properties in digital radiography. 6. MTFs of II–TV digital imaging systemsMedical Physics, 1985
- Investigation of basic imaging properties in digital radiography. 3. Effect of pixel size on SNR and threshold contrastMedical Physics, 1985