Cortical Screw Pullout Strength and Effective Shear Stress in Synthetic Third Generation Composite Femurs
Authors:
- 28 August 2006
- journal article
- research article
- Published by ASME International in Journal of Biomechanical Engineering
- Vol. 129 (2) , 289-293
Abstract
Background: The use of artificial bone analogs in biomechanical testing of orthopaedic fracture fixation devices has increased, particularly due to the recent development of commercially available femurs such as the third generation composite femur that closely reproduce the bulk mechanical behavior of human cadaveric and∕or fresh whole bone. The purpose of this investigation was to measure bone screw pullout forces in composite femurs and determine whether results are comparable to cadaver data from previous literature. Method of Approach: The pullout strengths of 3.5 and standard bicortical screws inserted into synthetic third generation composite femurs were measured and compared to existing adult human cadaveric and animal data from the literature. Results: For screws, the measured extraction shear stress in synthetic femurs was in the range of adult human femurs and tibias . For screws, the measured values in synthetic femurs were also similar to adult human specimens . Synthetic femur results for extraction stress showed no statistically significant site-to-site effect for 3.5 and screws, with one exception. Overall, the screws showed statistically higher stress required for extraction than screws. Conclusions: The third generation composite femurs provide a satisfactory biomechanical analog to human long-bones at the screw-bone interface. However, it is not known whether these femurs perform similarly to human bone during physiological screw “toggling.”
Keywords
This publication has 25 references indexed in Scilit:
- The Biomechanics of Human Femurs in Axial and Torsional Loading: Comparison of Finite Element Analysis, Human Cadaveric Femurs, and Synthetic FemursJournal of Biomechanical Engineering, 2006
- Fixation of Periprosthetic Femoral Shaft Fractures Associated With Cemented Femoral StemsJournal of Orthopaedic Trauma, 2006
- Unstable Proximal Extraarticular Tibia FracturesJournal of Orthopaedic Trauma, 2004
- Factors Affecting the Pullout Strength of Cancellous Bone ScrewsJournal of Biomechanical Engineering, 1996
- Mechanical validation of whole bone composite femur modelsJournal of Biomechanics, 1996
- Pullout strength of fixation screws from polymethylmethacrylate bone cementBiomaterials, 1995
- Caudo-Cephalad Loading of Pedicle ScrewsSpine, 1993
- Role of loads and prosthesis material properties on the mechanics of the proximal femur after total hip arthroplastyJournal of Orthopaedic Research, 1992
- The Holding Power of Orthopedic Screws In VivoClinical Orthopaedics and Related Research, 1975
- A study of some factors which affect the strength of screws and their insertion and holding power in boneJournal of Biomechanics, 1968