Two-dimensional stitching interferometry for self-calibration of high-order additive systematic errors
- 11 September 2019
- journal article
- research article
- Published by Optica Publishing Group in Optics Express
- Vol. 27 (19) , 26940-26956
Abstract
Stitching interferometry is performed by collecting interferometric data from overlapped sub-apertures and stitching these data together to provide a full surface map. The propagation of the systematic error in the measured subset data is one of the main error sources in stitching interferometry for accurate reconstruction of the surface topography. In this work, we propose, using the redundancy of the captured subset data, two types of two-dimensional (2D) self-calibration stitching algorithms to overcome this issue by in situ estimating the repeatable high-order additive systematic errors, especially for the application of measuring X-ray mirrors. The first algorithm, called CS short for “Calibrate, and then Stitch”, calibrates the high-order terms of the reference by minimizing the de-tilted discrepancies of the overlapped subsets and then stitches the reference-subtracted subsets. The second algorithm, called SC short for “Stitch, and then Calibrate”, stitches a temporarily result and then calibrates the reference from the de-tilted discrepancies of the measured subsets and the temporarily stitched result. In the implementation of 2D scans in $x$- and $y$-directions, step randomization is introduced to generate nonuniformly spaced subsets which can diminish the periodic stitching errors commonly observed in evenly spaced subsets. The regularization on low-order terms enables a highly flexible option to add the curvature and twist acquired by another system. Both numerical simulations and experiments are carried out to verify the proposed method. All the results indicate that 2D high-order repeatable additive systematic errors can be retrieved from the 2D redundant overlapped data in stitching interferometry.Keywords
Funding Information
- Office of Science (DE-SC0012704)
This publication has 21 references indexed in Scilit:
- Two-dimensional stitching interferometry based on tilt measurementOptics Express, 2018
- One-dimensional angular-measurement-based stitching interferometryOptics Express, 2018
- Stitching methods at the European Synchrotron Radiation Facility (ESRF)Review of Scientific Instruments, 2016
- Principles of interference microscopy for the measurement of surface topographyAdvances in Optics and Photonics, 2015
- Subaperture stitching interferometry of high-departure aspheres by incorporating a variable optical nullCIRP Annals, 2010
- Iterative algorithm for subaperture stitching interferometry for general surfacesJournal of the Optical Society of America A, 2005
- Relative angle determinable stitching interferometry for hard x-ray reflective opticsReview of Scientific Instruments, 2005
- Stitching Interferometry: A Flexible Solution for Surface MetrologyOptics and Photonics News, 2003
- Self-Calibration: Reversal, Redundancy, Error Separation, and ‘Absolute Testing’CIRP Annals, 1996
- High-speed noncontact profiler based on scanning white-light interferometryApplied Optics, 1994