A COMPARATIVE ANALYSIS OF COREGISTERED ULTRASOUND AND MAGNETIC RESONANCE IMAGING IN NEUROSURGERY
Authors:
- Alex HartovAHAlex Hartov
- David W. RobertsDRDavid W. Roberts
- Keith D. PaulsenKPKeith D. Paulsen
- 1 March 2008
- journal article
- research article
- Published by Wolters Kluwer Health in Operative Neurosurgery
- Vol. 62 (3) , 91-101
Abstract
This work presents qualitative and quantitative side-by-side comparisons of oblique coregistered magnetic resonance imaging (MRI) scans and ultrasound images obtained during 35 neurosurgical procedures. Spatially registered series of ultrasound images were recorded for subsequent off-line evaluation and comparison with corresponding preoperative MRI studies. The degree of misalignment was reduced by reregistering the target volume directly with segmented features. The initial apparent spatial misalignment of the target volume after craniotomy ranged from 0.11 to 8.73 mm (mean, 4.01 mm). After reregistration, the mutual information in overlapping segmented features was increased, presumably evidence of a better alignment locally. Additionally, the degree of feature congruence, which was assessed quantitatively through a convex hull approximation, demonstrated that the ultrasound volume was consistently smaller than its MRI counterpart. Although intraoperative ultrasound tends to be difficult to interpret by itself, when accurately coregistered with preoperative MRI scans, its potential utility as a navigational guide is enhanced.Keywords
This publication has 35 references indexed in Scilit:
- Functional neuronavigation combined with intra-operative 3D ultrasound: Initial experiences during surgical resections close to eloquent brain areas and future directions in automatic brain shift compensation of preoperative dataActa Neurochirurgica, 2007
- Computer-assisted 3D ultrasound-guided neurosurgery: technological contributions, including multimodal registration and advanced display, demonstrating future perspectivesThe International Journal of Medical Robotics and Computer Assisted Surgery, 2006
- Placement of deep brain stimulator electrodes using real‐time high‐field interventional magnetic resonance imagingMagnetic Resonance in Medicine, 2005
- Evaluation of intra-operative ultrasound imaging in brain tumor resection: a prospective studyNeurological Research, 2005
- Intraoperative landmarking of vascular anatomy by integration of duplex and Doppler ultrasonography in image-guided surgery. Technical noteSurgical Neurology, 2005
- Future perspectives for intraoperative MRINeurosurgery Clinics of North America, 2005
- Image fusion of MR images and real-time ultrasonography: evaluation of fusion accuracy combining two commercial instruments, a neuronavigation system and a ultrasound systemActa Neurochirurgica, 2004
- Quantification of, Visualization of, and Compensation for Brain Shift Using Intraoperative Magnetic Resonance ImagingNeurosurgery, 2000
- A comparison between preoperative magnetic resonance and intraoperative ultrasound tumor volumes and marginsJournal of Clinical Ultrasound, 1994
- Intraoperative ultrasound (US) imaging. Comparison of pathomorphological findings in US and CTActa Neurochirurgica, 1990