Modern Metal-on-Metal Hip Implants
Authors:
- James BrowneJBJames Browne
- Chris J. DanglesCDChris J. Dangles
- Paul A. MannerPMPaul A. Manner
- Adolph J. YatesAYAdolph J. Yates
- Kristy L. WeberKWKristy L. Weber
- Kevin M. BoyerKBKevin M. Boyer
- Paul ZemaitisPZPaul Zemaitis
- Anne WoznicaAWAnne Woznica
- Charles M. TurkelsonCTCharles M. Turkelson
- Janet L. WiesJWJanet L. Wies
- 1 June 2012
- journal article
- review article
- Published by Wolters Kluwer Health in Journal of the American Academy of Orthopaedic Surgeons
- Vol. 20 (6) , 402-406
Abstract
This Technology Overview was prepared using systematic review methodology and summarizes the findings of studies published as of July 15, 2011, on modern metal-on-metal hip implants. Analyses conducted on outcomes by two joint registries indicate that patients who receive metal-on-metal total hip arthroplasty (THA) and hip resurfacing are at greater risk for revision than are patients who receive THA using a different bearing surface combination. Data from these registries also indicate that larger femoral head components have higher revision rates and risk of revision and that older age is associated with increased revision risks of large-head metal-on-metal THA. Several studies noted a correlation between suboptimal hip implant positioning and higher wear rates, local metal debris release, and consequent local tissue reactions to metal debris. In addition, several studies reported elevated serum metal ion concentrations in patients with metal-on-metal hip articulations, although the clinical significance of these elevated ion concentrations remains unknown.Keywords
This publication has 18 references indexed in Scilit:
- Insufficient Acetabular Version Increases Blood Metal Ion Levels after Metal-on-metal Hip ResurfacingClinical Orthopaedics and Related Research, 2011
- Perivascular Lymphocytic Infiltration Is Not Limited to Metal-on-Metal BearingsClinical Orthopaedics and Related Research, 2011
- Elevation of circulating HLA DR+CD8+T-cells and correlation with chromium and cobalt concentrations 6 years after metal-on-metal hip arthroplastyActa Orthopaedica, 2010
- No Superiority of Cemented Metal-on-Metal Over Metal-on-Polyethylene THA in a Randomized Controlled Trial at 10-Year Follow-upOrthopedics, 2010
- Early failure of metal-on-metal bearings in hip resurfacing and large-diameter total hip replacementThe Journal of Bone and Joint Surgery. British volume, 2010
- Risk factors for inflammatory pseudotumour formation following hip resurfacingThe Journal of Bone and Joint Surgery. British volume, 2009
- The painful metal-on-metal hip resurfacingThe Journal of Bone and Joint Surgery. British volume, 2009
- Outcome and serum ion determination up to 11 years after implantation of a cemented metal-on-metal hip prosthesisActa Orthopaedica, 2009
- Does Ion Release Differ Between Hip Resurfacing and Metal-on-metal THA?Clinical Orthopaedics and Related Research, 2008