Real time tracking in liver SBRT: comparison of CyberKnife and Vero by planning structure-basedγ-evaluation and dose-area-histograms
- 2 February 2016
- journal article
- research article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 61 (4) , 1677-1691
Abstract
The purpose of this study was to evaluate and compare two clinical tracking systems for radiosurgery with regard to their dosimetric and geometrical accuracy in liver SBRT: the robot-based CyberKnife and the gimbal-based Vero. Both systems perform real-time tumour tracking by correlating internal tumour and external surrogate motion. CyberKnife treatment plans were delivered to a high resolution 2D detector array mounted on a 4D motion platform, with the platform simulating (a) tumour motion trajectories extracted from the corresponding CyberKnife predictor log files and (b) the tumour motion trajectories with superimposed baseline-drift. Static reference and tracked dose measurements were compared and dosimetric as well as geometrical uncertainties analyzed by a planning structure-based evaluation. For (a), γ -passing rates inside the CTV ( γ -criteria of 1% / 1 mm) ranged from 95% to 100% (CyberKnife) and 98% to 100% (Vero). However, dosimetric accuracy decreases in the presence of the baseline-drift. γ -passing rates for (b) ranged from 26% to 92% and 94% to 99%, respectively; i.e. the effect was more pronounced for CyberKnife. In contrast, the Vero system led to maximum dose deviations in the OAR between +1.5 Gy to +6.0 Gy (CyberKnife: +0.5 Gy to +3.5 Gy). Potential dose shifts were interpreted as motion-induced geometrical tracking errors. Maximum observed shift ranges were −1.0 mm to +0.7 mm (lateral) /−0.6 mm to +0.1 mm (superior–inferior) for CyberKnife and −0.8 mm to +0.2 mm /−0.8 mm to +0.4 mm for Vero. These values illustrate that CyberKnife and Vero provide high precision tracking of regular breathing patterns. Even for the modified motion trajectory, the obtained dose distributions appear to be clinical acceptable with regard to literature QA γ -criteria of 3% / 3 mm.Keywords
Funding Information
- Forschungsförderfond of the Medical Faculty of the University of Hamburg (FFM)
This publication has 22 references indexed in Scilit:
- Geometric Verification of Dynamic Wave Arc Delivery With the Vero System Using Orthogonal X-ray Fluoroscopic ImagingInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2015
- Design, performance characteristics and application examples of a new 4D motion platformZeitschrift für Medizinische Physik, 2015
- Initial assessment of tumor tracking with a gimbaled linac system in clinical circumstances: A patient simulation studyRadiotherapy and Oncology, 2013
- Correlation between external and internal respiratory motion: a validation studyInternational Journal of Computer Assisted Radiology and Surgery, 2011
- Report of AAPM TG 135: Quality assurance for robotic radiosurgeryMedical Physics, 2011
- Geometric accuracy of a novel gimbals based radiation therapy tumor tracking systemRadiotherapy and Oncology, 2011
- The CyberKnife® Robotic Radiosurgery System in 2010Technology in Cancer Research & Treatment, 2010
- The design, physical properties and clinical utility of an iris collimator for robotic radiosurgeryPhysics in Medicine & Biology, 2009
- Clinical Accuracy of the Respiratory Tumor Tracking System of the CyberKnife: Assessment by Analysis of Log FilesInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2009
- The management of respiratory motion in radiation oncology report of AAPM Task Group 76a)Medical Physics, 2006