Concordance between the new modular slit-lamp mounted Goldmann applanation tonometer (GAT), conventional slit-lamp mounted GAT, and Perkins tonometer
Authors:
Open Access
- 24 November 2025
- journal article
- research article
- Published by Wolters Kluwer Health in Indian Journal Of Ophthalmology
- Vol. 73 (12) , 1786-1791
Abstract
Purpose: To evaluate a new device, the Goldmann applanation tonometer (GAT) mounted on a novel handheld 3D-printed modular slit lamp (MSL), and compare it with the gold standard slit-lamp-mounted GAT and the handheld Perkins tonometer. Methods: One hundred eyes from 100 participants were included in this cross-sectional observational study. The MSL mount was customized for the selected tonometer (Model AT030, Carl Zeiss, India). A Samsung Galaxy S2 mobile phone was used for observing mires and photography during tonometry. One observer measured intraocular pressure (IOP) with slit-lamp GAT, and another observer measured with MSL GAT and Perkins tonometers. The tonometry sequence was slit-lamp GAT, MSL GAT, and Perkins tonometer, with a 5-minute interval between readings. Results: The median age of participants was 56 years (range: 16–83). The median IOP by slit-lamp GAT was 15 mmHg (range: 9–41). Tonometers were compared in pairs: slit-lamp GAT and Perkins, slit-lamp GAT and MSL GAT, and Perkins and MSL GAT. The pairwise Spearman rank correlations were 0.988, 0.988, and 0.983 (all P < 0.001). The mean differences in IOP (95% limits of agreement) were 0.5 (-0.6 to 1.6), 1.2 (0 to 2.4), and 0.7 (-0.7 to 2.1) mmHg, respectively. Conclusion: The MSL-mounted GAT provides reliable IOP readings with minor underestimation (within 2 mmHg in 88% of cases), suggesting its potential utility in clinical practice as a portable slit lamp and anterior segment imaging device.Keywords
This publication has 11 references indexed in Scilit:
- 3D printing in Ophthalmology: From medical implants to personalised medicineInternational Journal of Pharmaceutics, 2022
- Evaluating Goldmann Applanation Tonometry Intraocular Pressure Measurement Agreement Between Ophthalmic Technicians and PhysiciansAmerican Journal of Ophthalmology, 2020
- The Outcomes of a Comprehensive Program for Maintenance of Goldmann Applanation TonometerJournal of Glaucoma, 2019
- Applanation tonometry: a comparison of the Perkins handheld and Goldmann slit lamp-mounted methodsClinical Ophthalmology, 2014
- Intraocular Pressure Measurement Precision with the Goldmann Applanation, Dynamic Contour, and Ocular Response Analyzer TonometersOphthalmology, 2010
- Bar charts enhance Bland–Altman plots when value ranges are limitedJournal of Clinical Epidemiology, 2010
- Modalities of Tonometry and their Accuracy with Respect to Corneal Thickness and IrregularitiesJournal of Optometry, 2008
- Accurate time interval in repeated tonometryActa Ophthalmologica Scandinavica, 1998
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986
- Perkins hand-held applanation tonometer. A clinical evaluation.British Journal of Ophthalmology, 1968