Body‐loop related MRI radiofrequency‐induced heating hazards: Observations, characterizations, and recommendations
- 6 August 2021
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 87 (1) , 337-348
Abstract
Purpose: To assess RF‐induced heating hazards in 1.5T MR systems caused by body‐loop postures.Methods: Twelve advanced high‐resolution anatomically correct human body models with different body‐loop postures are created based on poseable human adult male models. Numerical simulations are performed to assess the radiofrequency (RF)‐induced heating of these 12 models at 11 landmarks. A customized phantom is developed to validate the numerical simulations and quantitatively analyze factors affecting the RF‐induced heating, eg, the contact area, the loop size, and the loading position. The RF‐induced heating inside three differently posed phantoms is measured.Results: The RF‐induced heating from the body‐loop postures can be up to 11 times higher than that from the original posture. The RF‐induced heating increases with increasing body‐loop size and decreasing contact area. The magnetic flux increases when the body‐loop center and the RF coil isocenter are close to each other, leading to increased RF‐induced heating. An air gap created in the body loop or generating a polarized magnetic field parallel to the body loop can reduce the heating by a factor of three at least. Experimental measurements are provided, validating the correctness of the numerical results.Conclusion: Safe patient posture during MR examinations is recommended with the use of insulation materials to prevent loop formation and consequently avoiding high RF‐induced heating. If body loops cannot be avoided, the body loop should be placed outside the RF transmitting coil. In addition, linear polarization with magnetic fields parallel to the body loop can be used to circumvent high RF‐induced heating.Keywords
Funding Information
- Division of Industrial Innovation and Partnerships (1922389)
This publication has 27 references indexed in Scilit:
- Whole‐body and local RF absorption in human models as a function of anatomy and position within 1.5T MR body coilMagnetic Resonance in Medicine, 2013
- The Virtual Family—development of surface-based anatomical models of two adults and two children for dosimetric simulationsPhysics in Medicine & Biology, 2009
- Quantitative analysis of the velocity and synchronicity of diaphragmatic motion: dynamic MRI in different posturesMagnetic Resonance Imaging, 2006
- Correlation Between Maximum Temperature Increase and Peak SAR with Different Average Schemes and MassesIEEE Transactions on Electromagnetic Compatibility, 2006
- Quantifying the effect of posture on intracranial physiology in humans by MRI flow studiesJournal of Magnetic Resonance Imaging, 2005
- The flexion gap in normal kneesThe Journal of Bone and Joint Surgery. British volume, 2004
- Evaluation of specific absorption rate as a dosimeter of MRI‐related implant heatingJournal of Magnetic Resonance Imaging, 2004
- Thermal Injuries Associated with MRIClinical Radiology, 2001
- Mechanism of burn injury during magnetic resonance imaging (MRI) — simple loops can induce heat injuryFrontiers of Medical and Biological Engineering, 2001
- Unusual burns of the lower extremities caused by a closed conducting loop in a patient at MR imaging.Radiology, 1996