Abstract
The complaint from hearing aid wearers of hollowness in the sound of their voice is typically associated with excessive low-frequency sound pressure level (SPL) in the ear canal. Increasing the vent diameter and/or reducing the gain in the low frequency would typically minimize this complaint. This paper reports on a case where the origin of hollowness was insufficient low-frequency gain compared to a previous hearing aid fitting. It describes the systematic process that was followed in uncovering the origin of the patient's hollowness complaint. Clinicians might follow a similar objective approach in their fine-tuning process to resolve wearer complaints. Spanish La queja de los usuarios sobre la cualidad de voz hueca que se percibe con el uso de sus auxiliares auditivos está típicamente asociada a un nivel excesivo de presión sonora (SPL) en las frecuencias graves dentro del canal auditivo. El incremento en el diámetro del orificio de ventilación y/o la reducción de la ganancia en las bajas frecuencias consigue minimizar esta queja. Este artículo reporta un caso donde el origen de esta característica en la voz fue una ganancia insuficiente en las frecuencias graves, en comparación con el auxiliar auditivo previo. Describe el proceso sistemático que se siguió para descubrir el origen de la queja del paciente sobre la apreciación de oquedad en su voz. Los clínicos podrían seguir un enfoque objetivo similar en su proceso de resolver las quejas de los usuarios.