Francis Kuk
Research Subjects
Research Interests
Francis Kuk has a total of 14 co-authors. Here are the top co-authors by number of shared publications:
Denise Keenan
8 shared publicationsChi-Chuen Lau
6 shared publicationsCarl Ludvigsen
2 shared publicationsPetri Korhonen
2 shared publicationsAmy Jackson
1 shared publicationCarsten Paludan-Müller
Francis K. Kuk, PhD, is Director of Audiology, Widex USA. Carsten Paludan-Müller, MSc, is a Research Engineer at Widex A/S
1 shared publication
Latest Publications
- 1 September 2009
- Published by American Academy of Audiology in Journal of the American Academy of Audiology
- Vol. 20 (8) , 465-479
Efficacy of Linear Frequency Transposition on Consonant Identification in Quiet and in Noise
Background: Frequency transposition has gained renewed interest in recent years. This type of processing takes sounds in the unaidable high-frequency region and moves them to the lower frequency region. One concern is that the transposed sounds mask or distort the original low-frequency sounds and lead to a poorer performance. On the other hand, experience with transposition may allow the listeners to relearn the new auditory percepts and benefit from transposition. Purpose: The current study was designed to examine the effect of linear frequency transposition on consonant identification in quiet (50 dB SPL and 68 dB SPL) and in noise at three intervals—the initial fit, after one month of use (along with auditory training), and a further one month of use (without directed training) of transposition. Research Design: A single-blind, factorial repeated-measures design was used to study the effect of test conditions (three) and hearing aid setting/time interval (four) on consonant identification. Study Sample: Eight adults with a severe-to-profound high-frequency sensorineural hearing loss participated. Intervention: Participants were fit with the Widex m4-m behind-the-ear hearing aids binaurally in the frequency transposition mode, and their speech scores were measured initially. They wore the hearing aids home for one month and were instructed to complete a self-paced “bottom-up” training regimen. They returned after the training, and their speech performance was measured. They wore the hearing aids home for another month, but they were not instructed to complete any auditory training. Their speech performance was again measured at the end of the two-month trial. Data Collection and Analysis: Consonant performance was measured with a nonsense syllable test (ORCA-NST) that was developed at this facility (Office of Research in Clinical Amplification [Widex]). The test conditions included testing in quiet at 50 dB SPL and 68 dB SPL, and at 68 dB SPL in noise (SNR [signal-to-noise ratio] = +5). The hearing aid conditions included no transposition at initial fit (V1), transposition at initial fit (V2), transposition at one month post-fit (V3), and transposition at 2 months post-fit (V4). Identification scores were analyzed for each individual phoneme and phonemic class. Repeated-measures ANOVA were conducted using SPSS software to examine significant differences. Results: For all test conditions (50 dB SPL in quiet, 68 dB SPL in quiet, and 68 dB SPL in noise), a statistically significant difference (p < 0.05 level) was reached between the transposition condition measured at two months postfitting and the initial fitting (with and without transposition) for fricatives only. The difference between transposition and the no-transposition conditions at the 50 dB SPL condition was also significant for the initial and one-month intervals. Analysis of individual phonemes showed a decrease in the number of confusions and an increase in the number of correct identification over time. Conclusions: Linear frequency transposition improved fricative identification over time. Proper candidate selection with appropriate training is necessary to fully realize the potential benefit of this type of processing.
- 1 September 2009
- Published by American Academy of Audiology in Journal of the American Academy of Audiology
- Vol. 20 (8) , 465-479
Efficacy of Linear Frequency Transposition on Consonant Identification in Quiet and in Noise
Background: Frequency transposition has gained renewed interest in recent years. This type of processing takes sounds in the unaidable high-frequency region and moves them to the lower frequency region. One concern is that the transposed sounds mask or distort the original low-frequency sounds and lead to a poorer performance. On the other hand, experience with transposition may allow the listeners to relearn the new auditory percepts and benefit from transposition. Purpose: The current study was designed to examine the effect of linear frequency transposition on consonant identification in quiet (50 dB SPL and 68 dB SPL) and in noise at three intervals—the initial fit, after one month of use (along with auditory training), and a further one month of use (without directed training) of transposition. Research Design: A single-blind, factorial repeated-measures design was used to study the effect of test conditions (three) and hearing aid setting/time interval (four) on consonant identification. Study Sample: Eight adults with a severe-to-profound high-frequency sensorineural hearing loss participated. Intervention: Participants were fit with the Widex m4-m behind-the-ear hearing aids binaurally in the frequency transposition mode, and their speech scores were measured initially. They wore the hearing aids home for one month and were instructed to complete a self-paced “bottom-up” training regimen. They returned after the training, and their speech performance was measured. They wore the hearing aids home for another month, but they were not instructed to complete any auditory training. Their speech performance was again measured at the end of the two-month trial. Data Collection and Analysis: Consonant performance was measured with a nonsense syllable test (ORCA-NST) that was developed at this facility (Office of Research in Clinical Amplification [Widex]). The test conditions included testing in quiet at 50 dB SPL and 68 dB SPL, and at 68 dB SPL in noise (SNR [signal-to-noise ratio] = +5). The hearing aid conditions included no transposition at initial fit (V1), transposition at initial fit (V2), transposition at one month post-fit (V3), and transposition at 2 months post-fit (V4). Identification scores were analyzed for each individual phoneme and phonemic class. Repeated-measures ANOVA were conducted using SPSS software to examine significant differences. Results: For all test conditions (50 dB SPL in quiet, 68 dB SPL in quiet, and 68 dB SPL in noise), a statistically significant difference (p < 0.05 level) was reached between the transposition condition measured at two months postfitting and the initial fitting (with and without transposition) for fricatives only. The difference between transposition and the no-transposition conditions at the 50 dB SPL condition was also significant for the initial and one-month intervals. Analysis of individual phonemes showed a decrease in the number of confusions and an increase in the number of correct identification over time. Conclusions: Linear frequency transposition improved fricative identification over time. Proper candidate selection with appropriate training is necessary to fully realize the potential benefit of this type of processing.
- 1 September 2009
- Published by American Academy of Audiology in Journal of the American Academy of Audiology
- Vol. 20 (8) , 480-491
Comparison of Vent Effects between a Solid Earmold and a Hollow Earmold
Background: Hollow earmolds have become a popular type of earmold used in thin-tube, microsize hearing aid fittings. It is desirable for clinicians to be familiar with their characteristics and limitations. Purpose: This investigation compared the effects of vent diameter between a traditional solid earmold and a hollow earmold that is used in modern thin-tube hearing aid fittings. Research Design: A single-blind, 2 × 4 factorial design was used. Study Sample: Eight adults with a high-frequency hearing loss participated. Intervention: Custom earmolds for use with thin-tube hearing aids were made for each participant. Two types of earmolds were made: a solid earmold with a traditional vent length and a hollow earmold where the thickness of the shell was the length of the vent. Vent diameters were 0, 1, 2, and 3 mm. Data Collection and Analysis: The vent effect was evaluated on real-ear aided response, real-ear occluded response during vocalization, subjective occlusion rating, insertion loss, and maximum available gain before feedback. Real-ear measurements were made with the Fonix 6500 probe-microphone real-ear system. Vocalizations from the participants were analyzed with a custom MATLAB program, and statistical analysis was conducted with SPSS software. Results: A systematic vent effect was seen with each earmold type as the nominal vent diameter changed. For the same vent diameter, the vent effect seen with the hollow earmold was greater than that of the solid earmold. Conclusions: Because of the difference in vent length (and thus acoustic mass) between a solid and a hollow earmold, the effect of vent diameter in a hollow earmold is more pronounced than that seen in a solid earmold of the same nominal vent diameter. Thus, a smaller vent diameter will be needed in a hollow earmold than in a solid earmold to achieve similar vent effects.
- 1 September 2009
- Published by American Academy of Audiology in Journal of the American Academy of Audiology
- Vol. 20 (8) , 480-491
Comparison of Vent Effects between a Solid Earmold and a Hollow Earmold
Background: Hollow earmolds have become a popular type of earmold used in thin-tube, microsize hearing aid fittings. It is desirable for clinicians to be familiar with their characteristics and limitations. Purpose: This investigation compared the effects of vent diameter between a traditional solid earmold and a hollow earmold that is used in modern thin-tube hearing aid fittings. Research Design: A single-blind, 2 × 4 factorial design was used. Study Sample: Eight adults with a high-frequency hearing loss participated. Intervention: Custom earmolds for use with thin-tube hearing aids were made for each participant. Two types of earmolds were made: a solid earmold with a traditional vent length and a hollow earmold where the thickness of the shell was the length of the vent. Vent diameters were 0, 1, 2, and 3 mm. Data Collection and Analysis: The vent effect was evaluated on real-ear aided response, real-ear occluded response during vocalization, subjective occlusion rating, insertion loss, and maximum available gain before feedback. Real-ear measurements were made with the Fonix 6500 probe-microphone real-ear system. Vocalizations from the participants were analyzed with a custom MATLAB program, and statistical analysis was conducted with SPSS software. Results: A systematic vent effect was seen with each earmold type as the nominal vent diameter changed. For the same vent diameter, the vent effect seen with the hollow earmold was greater than that of the solid earmold. Conclusions: Because of the difference in vent length (and thus acoustic mass) between a solid and a hollow earmold, the effect of vent diameter in a hollow earmold is more pronounced than that seen in a solid earmold of the same nominal vent diameter. Thus, a smaller vent diameter will be needed in a hollow earmold than in a solid earmold to achieve similar vent effects.
- 1 January 2009
- Published by American Speech Language Hearing Association in The ASHA Leader
- Vol. 14 (1) , 14-17
Re-evaluating the Efficacy of Frequency Transposition
Patients with a precipitous high-frequency hearing loss pose a unique challenge for dispensing audiologists. Hearing aids rarely provide sufficient high-frequency gain without over-amplifying the lower frequencies, which results in poor sound quality or significant sound distortions. The occlusion effect and upward spread of masking may also occur. For these patients, the high frequencies are not easily reachable.
- 1 January 2009
- Published by American Speech Language Hearing Association in The ASHA Leader
- Vol. 14 (1) , 14-17
Re-evaluating the Efficacy of Frequency Transposition
Patients with a precipitous high-frequency hearing loss pose a unique challenge for dispensing audiologists. Hearing aids rarely provide sufficient high-frequency gain without over-amplifying the lower frequencies, which results in poor sound quality or significant sound distortions. The occlusion effect and upward spread of masking may also occur. For these patients, the high frequencies are not easily reachable.
- 1 June 2008
- Published by American Academy of Audiology in Journal of the American Academy of Audiology
- Vol. 19 (6) , 465-480
Personal Amplification for School-Age Children with Auditory Processing Disorders
Background: Children with auditory processing disorders (APD) are described to have a signal-to-noise ratio (SNR) difficulty. Frequency-modulated (FM) systems have been reported to improve this situation. Yet the use of personal amplification that may be more portable has not been attempted. Purpose: To determine whether personal amplification would result in improvement in speech-in-noise performances (attentiveness and speech recognition) and daily functioning in children with diagnosed APD. In addition, the desired hearing aid features (such as required gain, directional microphone and noise reduction, and open-ear fitting) are examined. Research Design: A single-blind, longitudinal descriptive study in which subjects served as their own control in various hearing aid conditions. Study Sample: Fourteen normal hearing children who had a diagnosis of APD and who were between the ages of 7 and 11 participated. Intervention: All subjects wore bilateral, mild-gain, behind-the-ear, wide dynamic range compression hearing aids fitted in an open-ear mode. Gain on the hearing aids was adjusted to provide approximately 10 dB of insertion gain for conversational input. Directional microphone and noise reduction were used on the hearing aids. Subjects wore the hearing aids home and were encouraged to use them as much as possible in their daily environments (school, home, and social activities). Subjects were seen four times: an initial visit where hearing aids were fitted, then visits at 2 weeks, 3 months, and 6 months after the initial fitting. The majority of the testing was completed during these final three visits. Data Collection and Analysis: The children were evaluated on the Northwestern University word-list (NU‐6) and the Auditory Continuous Performance Test (ACPT) in noise at most visits. The hearing aids were evaluated in the omnidirectional microphone mode only, omnidirectional microphone with noise reduction (NR) mode, and directional microphone with NR mode. The children's parents and teachers were asked to complete the Children's Auditory Processing Performance Scale (CHAPS) questionnaire both before and at the end of the study. Results: The results showed that the use of hearing aids in the omnidirectional microphone mode alone did not improve speech identification in noise over the unaided condition. However, the inclusion of the NR algorithm and directional microphones improved speech understanding in noise. Amplification reduced the number of errors on the ACPT and improved several areas on the CHAPS; however, the results were not statistically significant. Conclusions: The use of mild-gain, open-ear fitting hearing aids with a directional microphone and noise reduction algorithm may be attempted on some children with APD on a trial basis.
- 1 June 2008
- Published by American Academy of Audiology in Journal of the American Academy of Audiology
- Vol. 19 (6) , 465-480
Personal Amplification for School-Age Children with Auditory Processing Disorders
Background: Children with auditory processing disorders (APD) are described to have a signal-to-noise ratio (SNR) difficulty. Frequency-modulated (FM) systems have been reported to improve this situation. Yet the use of personal amplification that may be more portable has not been attempted. Purpose: To determine whether personal amplification would result in improvement in speech-in-noise performances (attentiveness and speech recognition) and daily functioning in children with diagnosed APD. In addition, the desired hearing aid features (such as required gain, directional microphone and noise reduction, and open-ear fitting) are examined. Research Design: A single-blind, longitudinal descriptive study in which subjects served as their own control in various hearing aid conditions. Study Sample: Fourteen normal hearing children who had a diagnosis of APD and who were between the ages of 7 and 11 participated. Intervention: All subjects wore bilateral, mild-gain, behind-the-ear, wide dynamic range compression hearing aids fitted in an open-ear mode. Gain on the hearing aids was adjusted to provide approximately 10 dB of insertion gain for conversational input. Directional microphone and noise reduction were used on the hearing aids. Subjects wore the hearing aids home and were encouraged to use them as much as possible in their daily environments (school, home, and social activities). Subjects were seen four times: an initial visit where hearing aids were fitted, then visits at 2 weeks, 3 months, and 6 months after the initial fitting. The majority of the testing was completed during these final three visits. Data Collection and Analysis: The children were evaluated on the Northwestern University word-list (NU‐6) and the Auditory Continuous Performance Test (ACPT) in noise at most visits. The hearing aids were evaluated in the omnidirectional microphone mode only, omnidirectional microphone with noise reduction (NR) mode, and directional microphone with NR mode. The children's parents and teachers were asked to complete the Children's Auditory Processing Performance Scale (CHAPS) questionnaire both before and at the end of the study. Results: The results showed that the use of hearing aids in the omnidirectional microphone mode alone did not improve speech identification in noise over the unaided condition. However, the inclusion of the NR algorithm and directional microphones improved speech understanding in noise. Amplification reduced the number of errors on the ACPT and improved several areas on the CHAPS; however, the results were not statistically significant. Conclusions: The use of mild-gain, open-ear fitting hearing aids with a directional microphone and noise reduction algorithm may be attempted on some children with APD on a trial basis.
- 1 April 2008
- Published by Wolters Kluwer Health in The Hearing Journal
- Vol. 61 (4) , 50
Criteria for evaluating the performance of linear frequency transposition in children
- 1 April 2008
- Published by Wolters Kluwer Health in The Hearing Journal
- Vol. 61 (4) , 50
Criteria for evaluating the performance of linear frequency transposition in children
- Open Access
- 1 April 2007
- Published by Wolters Kluwer Health in The Hearing Journal
- Vol. 60 (4) , 59-63
Use of frequency transposition in a thin-tube open-ear fitting
- 1 April 2007
- Published by Wolters Kluwer Health in The Hearing Journal
- Vol. 60 (4) , 59-63
Open AccessUse of frequency transposition in a thin-tube open-ear fitting
- 1 April 2006
- Published by Wolters Kluwer Health in The Hearing Journal
- Vol. 59 (4) , 62
Noise-management algorithm may improve speech intelligibility in noise
- 1 April 2006
- Published by Wolters Kluwer Health in The Hearing Journal
- Vol. 59 (4) , 62
Noise-management algorithm may improve speech intelligibility in noise
- 1 November 2005
- Published by American Academy of Audiology in Journal of the American Academy of Audiology
- Vol. 16 (10) , 781-788
Managing an "Own Voice" Problem That Has an Amplifier Origin
The complaint from hearing aid wearers of hollowness in the sound of their voice is typically associated with excessive low-frequency sound pressure level (SPL) in the ear canal. Increasing the vent diameter and/or reducing the gain in the low frequency would typically minimize this complaint. This paper reports on a case where the origin of hollowness was insufficient low-frequency gain compared to a previous hearing aid fitting. It describes the systematic process that was followed in uncovering the origin of the patient's hollowness complaint. Clinicians might follow a similar objective approach in their fine-tuning process to resolve wearer complaints. Spanish La queja de los usuarios sobre la cualidad de voz hueca que se percibe con el uso de sus auxiliares auditivos está típicamente asociada a un nivel excesivo de presión sonora (SPL) en las frecuencias graves dentro del canal auditivo. El incremento en el diámetro del orificio de ventilación y/o la reducción de la ganancia en las bajas frecuencias consigue minimizar esta queja. Este artículo reporta un caso donde el origen de esta característica en la voz fue una ganancia insuficiente en las frecuencias graves, en comparación con el auxiliar auditivo previo. Describe el proceso sistemático que se siguió para descubrir el origen de la queja del paciente sobre la apreciación de oquedad en su voz. Los clínicos podrían seguir un enfoque objetivo similar en su proceso de resolver las quejas de los usuarios.
- 1 November 2005
- Published by American Academy of Audiology in Journal of the American Academy of Audiology
- Vol. 16 (10) , 781-788
Managing an "Own Voice" Problem That Has an Amplifier Origin
The complaint from hearing aid wearers of hollowness in the sound of their voice is typically associated with excessive low-frequency sound pressure level (SPL) in the ear canal. Increasing the vent diameter and/or reducing the gain in the low frequency would typically minimize this complaint. This paper reports on a case where the origin of hollowness was insufficient low-frequency gain compared to a previous hearing aid fitting. It describes the systematic process that was followed in uncovering the origin of the patient's hollowness complaint. Clinicians might follow a similar objective approach in their fine-tuning process to resolve wearer complaints. Spanish La queja de los usuarios sobre la cualidad de voz hueca que se percibe con el uso de sus auxiliares auditivos está típicamente asociada a un nivel excesivo de presión sonora (SPL) en las frecuencias graves dentro del canal auditivo. El incremento en el diámetro del orificio de ventilación y/o la reducción de la ganancia en las bajas frecuencias consigue minimizar esta queja. Este artículo reporta un caso donde el origen de esta característica en la voz fue una ganancia insuficiente en las frecuencias graves, en comparación con el auxiliar auditivo previo. Describe el proceso sistemático que se siguió para descubrir el origen de la queja del paciente sobre la apreciación de oquedad en su voz. Los clínicos podrían seguir un enfoque objetivo similar en su proceso de resolver las quejas de los usuarios.
- 1 October 2005
- Published by American Academy of Audiology in Journal of the American Academy of Audiology
- Vol. 16 (9) , 653-661
Real-World Performance of a Reverse-Horn Vent
The present study compared differences in subjective and objective performance in completely-in-the-canal (CIC) hearing aids with conventional uniform 1.5 mm parallel vents and another with a reverse horn vent where the diameter increased from 1.5 mm on the lateral faceplate to 3 mm on the medial opening of the hearing aid. Nine hearing-impaired persons with a high-frequency hearing loss participated. The test battery included unaided in situ thresholds, amount of available gain before feedback, speech in quiet, speech in noise (HINT), subjective ratings of hollowness and tolerance, objective measures of the occlusion effect, and real-ear aided response. Results showed less available gain before feedback but less occlusion effect for subjective ratings and objective measures with the reverse horn vent. This type of vent design may be useful to increase the effective vent diameter of custom (including CIC) hearing aids. Spanish El presente estudio compara las diferencia subjetivas y objetivas en el desempeño con auxiliares auditivos completamente dentro del canal (CIC) con orificios de ventilación paralelos, de 1.5 mm, uniformes y convencionales, y con orificio de ventilación de cuerno reverso, donde el diámetro se incrementa de 1.5 mm en la placa lateral a 3 mm en la apertura medial del auxiliar auditivo. Participaron nueve personas con hipoacusia en las altas frecuencias. La batería de pruebas incluyó umbrales in situ no amplificados, cantidad de ganancia disponible antes del inicio de la retroalimentación, lenguaje en silencio, lenguaje en ruido (HINT), apreciación subjetiva de sonido hueco y tolerancia, medidas objetivas del efecto de oclusión, y respuesta amplificada de oído real. Los resultados mostraron una menor ganancia disponible antes de la retroalimentación, pero un menor efecto de oclusión en las apreciaciones subjetivas y en las medidas objetivas con el orificio de ventilación de cuerno reverso. Este tipo de ventilación puede ser útil para incrementar el diámetro efectivo de dicha ventilación en auxiliares auditivos hechos a la medida (incluido el CIC).
- 1 October 2005
- Published by American Academy of Audiology in Journal of the American Academy of Audiology
- Vol. 16 (9) , 653-661
Real-World Performance of a Reverse-Horn Vent
The present study compared differences in subjective and objective performance in completely-in-the-canal (CIC) hearing aids with conventional uniform 1.5 mm parallel vents and another with a reverse horn vent where the diameter increased from 1.5 mm on the lateral faceplate to 3 mm on the medial opening of the hearing aid. Nine hearing-impaired persons with a high-frequency hearing loss participated. The test battery included unaided in situ thresholds, amount of available gain before feedback, speech in quiet, speech in noise (HINT), subjective ratings of hollowness and tolerance, objective measures of the occlusion effect, and real-ear aided response. Results showed less available gain before feedback but less occlusion effect for subjective ratings and objective measures with the reverse horn vent. This type of vent design may be useful to increase the effective vent diameter of custom (including CIC) hearing aids. Spanish El presente estudio compara las diferencia subjetivas y objetivas en el desempeño con auxiliares auditivos completamente dentro del canal (CIC) con orificios de ventilación paralelos, de 1.5 mm, uniformes y convencionales, y con orificio de ventilación de cuerno reverso, donde el diámetro se incrementa de 1.5 mm en la placa lateral a 3 mm en la apertura medial del auxiliar auditivo. Participaron nueve personas con hipoacusia en las altas frecuencias. La batería de pruebas incluyó umbrales in situ no amplificados, cantidad de ganancia disponible antes del inicio de la retroalimentación, lenguaje en silencio, lenguaje en ruido (HINT), apreciación subjetiva de sonido hueco y tolerancia, medidas objetivas del efecto de oclusión, y respuesta amplificada de oído real. Los resultados mostraron una menor ganancia disponible antes de la retroalimentación, pero un menor efecto de oclusión en las apreciaciones subjetivas y en las medidas objetivas con el orificio de ventilación de cuerno reverso. Este tipo de ventilación puede ser útil para incrementar el diámetro efectivo de dicha ventilación en auxiliares auditivos hechos a la medida (incluido el CIC).
- 1 October 2005
- Published by American Academy of Audiology in Journal of the American Academy of Audiology
- Vol. 16 (9) , 747-762
Vent Configurations on Subjective and Objective Occlusion Effect
The current study reexamined the effect of vent diameters on objective and subjective occlusion effect (OE) while minimizing two possible sources of variability. Nine hearing-impaired participants with primarily a high-frequency hearing loss were evaluated. Laser shell-making technology was used to make ear inserts of completely-in-the-canal (CIC) hearing aids for the study. This was to minimize any potential slit leakage from the inserts. The vent dimensions were systematically altered during the study. Participants sustained /i/ for 5 sec, and the real-ear occluded response was measured with a custom-made program that performed frequency averaging to reduce response variability. Participants also repeated the phrase "Baby Jeannie is teeny tiny" and rated their own voice. The results showed a systematic change in the objective OE and subjective ratings of OE as the vent diameter was modified. Furthermore, a significant correlation was seen between subjective rating and objective occlusion effect. El actual estudio re-examinó el efecto de los diferentes diámetros en el orificio de ventilación sobre el efecto de oclusión (OE), objetiva y subjetivamente, minimizando dos posibles fuentes de variabilidad. Se evaluó a nueve participantes hipoacúsicos con pérdidas auditivas primariamente en las altas frecuencias. Se utilizó una tecnología laser de fabricación de conchas para producir, para el estudio, dispositivos de inserción de auxiliares auditivos completamente en el canal (CIC). Esto se hizo para minimizar cualquier fuga de hendidura de los insertos. Las dimensiones de orificio de ventilación fueron sistemáticamente alteradas durante el estudio. Los participantes produjeron una /i/ sostenida por 5 segundos, y se midieron las respuestas ocluidas de oído real con un programa hecho a la medida que realizaba promediaciones de la frecuencia para reducir la variabilidad en la respuesta. Los participantes también repitieron la frase "Baby Jeannie is teeny tiny" y calificaron su propia voz. Los resultados mostraron un cambio sistemático en las apreciaciones objetivas y subjetivas del OE conforme se modificó el tamaño del orificio de ventilación. Más aún, se vio una correlación significativa entre la apreciación subjetiva y el efecto objetivo de oclusión.
- 1 October 2005
- Published by American Academy of Audiology in Journal of the American Academy of Audiology
- Vol. 16 (9) , 747-762
Vent Configurations on Subjective and Objective Occlusion Effect
The current study reexamined the effect of vent diameters on objective and subjective occlusion effect (OE) while minimizing two possible sources of variability. Nine hearing-impaired participants with primarily a high-frequency hearing loss were evaluated. Laser shell-making technology was used to make ear inserts of completely-in-the-canal (CIC) hearing aids for the study. This was to minimize any potential slit leakage from the inserts. The vent dimensions were systematically altered during the study. Participants sustained /i/ for 5 sec, and the real-ear occluded response was measured with a custom-made program that performed frequency averaging to reduce response variability. Participants also repeated the phrase "Baby Jeannie is teeny tiny" and rated their own voice. The results showed a systematic change in the objective OE and subjective ratings of OE as the vent diameter was modified. Furthermore, a significant correlation was seen between subjective rating and objective occlusion effect. El actual estudio re-examinó el efecto de los diferentes diámetros en el orificio de ventilación sobre el efecto de oclusión (OE), objetiva y subjetivamente, minimizando dos posibles fuentes de variabilidad. Se evaluó a nueve participantes hipoacúsicos con pérdidas auditivas primariamente en las altas frecuencias. Se utilizó una tecnología laser de fabricación de conchas para producir, para el estudio, dispositivos de inserción de auxiliares auditivos completamente en el canal (CIC). Esto se hizo para minimizar cualquier fuga de hendidura de los insertos. Las dimensiones de orificio de ventilación fueron sistemáticamente alteradas durante el estudio. Los participantes produjeron una /i/ sostenida por 5 segundos, y se midieron las respuestas ocluidas de oído real con un programa hecho a la medida que realizaba promediaciones de la frecuencia para reducir la variabilidad en la respuesta. Los participantes también repitieron la frase "Baby Jeannie is teeny tiny" y calificaron su propia voz. Los resultados mostraron un cambio sistemático en las apreciaciones objetivas y subjetivas del OE conforme se modificó el tamaño del orificio de ventilación. Más aún, se vio una correlación significativa entre la apreciación subjetiva y el efecto objetivo de oclusión.