Evaluation of Torque Measurement Surrogates as Applied to Grip Torque and Jaw Angle Estimation of Robotic Surgical Tools
- 22 June 2018
- journal article
- research article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Robotics and Automation Letters
- Vol. 3 (4) , 3027-3034
Abstract
The estimation of grip force for surgical tools such as the da Vinci has been shown to be valuable in possible applications such as haptics, tissue identification, and surgical training. Successful estimation attempts have been previously demonstrated, but utilize customized sensors; this work aims to provide an estimate considering only typical sensor streams already present in commercially available surgical robots. The objective of this work is to evaluate three proximal-end torque surrogate methods in their abilities to estimate distal-end states. The estimates are compared with previously reported results found in literature and the percent difference between the customized sensor approach and previous standards is reported. The most effective surrogates for proximal-end torque were commanded motor current and measured motor current. The jaw angle estimate resulted in 0.37 degree root mean square error, and the distal-end torque estimate resulted in 4.42 mNm RMSE, which compares favorably to existing literature approaches.Keywords
Funding Information
- Army Research Laboratory and was accomplished under Cooperative Agreement (W911NF-14-2-0035)
- National Science Foundation Graduate Research Fellowship (00039202)
This publication has 14 references indexed in Scilit:
- Dynamic modeling of cable driven elongated surgical instruments for sensorless grip force estimation2016
- Heterogeneous Sensor Data Fusion Approach for Real-time Monitoring in Ultraprecision Machining (UPM) Process Using Non-Parametric Bayesian Clustering and Evidence TheoryIEEE Transactions on Automation Science and Engineering, 2015
- Precision Farming: Sensor AnalyticsIEEE Intelligent Systems, 2015
- Learning by observation for surgical subtasks: Multilateral cutting of 3D viscoelastic and 2D Orthotropic Tissue Phantoms2015
- A grip force model for the da Vinci end-effector to predict a compensation forceMedical & Biological Engineering & Computing, 2014
- Quasi-static modeling of the da Vinci instrument2014
- Autonomous multilateral debridement with the Raven surgical robot2014
- The value of haptic feedback in conventional and robot-assisted minimal invasive surgery and virtual reality training: a current reviewSurgical Endoscopy, 2009
- Haptic feedback in robot-assisted minimally invasive surgeryCurrent Opinion in Urology, 2009
- Gauss-Newton approximation to Bayesian learning2002