Abstract
Depth-of-interaction (DOI) positron emission tomography (PET) detectors are essential in high-resolution small animal PET. The dual-ended readout detector configuration provides high DOI resolution; however, signal multiplexing schemes are often needed to reduce manufacturing costs. TOFPET2 is a low-cost, multi-channel application-specific integrated circuit (ASIC) designed for individual silicon photomultiplier (SiPM) readout, with performance in multiplexed configurations not well known. This study evaluates a highly multiplexed, high-resolution, dual-ended readout PET detector using TOFPET2. The PET detector was constructed with a 23 × 23 array of LYSO crystals (0.785 × 0.785 × 20 mm) coupled to 6 × 6 SiPM arrays (pitch = 3.3 mm) via 1.2 mm PMMA sheets. Custom boards reduced 72 SiPM signals to 8 (4 per end) before feeding into TOFPET2 channels. The detector was biased at 27, 27.5, and 28 V, with ASIC discriminator thresholds optimised based on flood map quality. Optimal flood maps were obtained at 27.5 V. The mean energy and DOI resolution were 10.5 ± 1.6% and 2.3 ± 0.3 mm, respectively, with a coincidence time resolution (CTR) of 1.3 ns between two detectors. This work demonstrates the feasibility of employing TOFPET2 to achieve high spatial, energy, and DOI resolution in a 9:1 multiplexed system.
Funding Information
  • Australian Research Council (DP190102318)
  • Mexican National Council of Humanities, Sciences and Technologies

This publication has 58 references indexed in Scilit: