Polymer Gap Adapter for Contactless, Robust, and Fast Measurements at 220–325 GHz
- 3 December 2015
- journal article
- research article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in Journal of Microelectromechanical Systems
- Vol. 25 (1) , 160-169
Abstract
Radiation leakages are a considerable problem when measuring waveguide structures at high frequencies. In order to maintain good electrical contact, flanges need to be tightly and evenly screwed to the device under test. This can be a time-consuming operation, especially with repeated measurements. We present a metamaterial-based adapter, which prohibits leakage even in the presence of gaps at the interconnects. This so-called gap adapter has been fabricated from a metallized polymer (SU8). The reflection coefficient is below -20 dB throughout the band for a 50-μm gap on both sides of the gap adapter. In comparison, a conventional waveguide with a 50-μm gap on both sides has a reflection coefficient of -10 dB. The gap adapter can be used to perform fast measurements, since the normal flange screws are redundant. We compare the SU8 gap adapter with a Si version and to a smooth metal waveguide reference disc. The SU8 gap adapter performed better than the Si version and much better than the waveguide disc in all test cases. SU8 gap adapters were used to measure on a waveguide component. The SU8 gap adapters with 50-μm gaps performed comparable with the waveguide component with the flange screws carefully tightened. The polymer also makes the gap adapter mechanically robust and easy to mass fabricate.Keywords
Funding Information
- Vetenskapsrådet
- European Research Council (ERC-2012-ADG 20120216)
- Chalmers Production Area of Advanced Science
This publication has 22 references indexed in Scilit:
- Micromachined contactless pin-flange adapter for robust high-frequency measurementsJournal of Micromechanics and Microengineering, 2014
- SU8 ridge-gap waveguide resonatorInternational Journal of Microwave and Wireless Technologies, 2014
- Planar Dual-Mode Horn Array With Corporate-Feed Network in Inverted Microstrip Gap WaveguideIEEE Transactions on Antennas and Propagation, 2014
- Study of Q ‐factors of ridge and groove gap waveguide resonatorsIET Microwaves, Antennas & Propagation, 2013
- An innovative waveguide interface and quarter-wavelength shim for the 220∓325 GHz band2012
- Improved Microstrip Filters Using PMC Packaging by Lid of NailsIEEE Transactions on Components, Packaging and Manufacturing Technology, 2012
- Contactless non-leaking waveguide flange realized by bed of nails for millimeter wave applications2012
- Measurements of micromachined waveguide devices at WR-3 band using a T/R-T module based network analyzer2011
- Numerical studies of bandwidth of parallel-plate cut-off realised by a bed of nails, corrugations and mushroom-type electromagnetic bandgap for use in gap waveguidesIET Microwaves, Antennas & Propagation, 2011
- Millimeter Wave Vector Analysis Calibration and Measurement Problems Caused by Common Waveguide Irregularities2000