Minimum and optimum light requirements for laryngoscopy in paediatric anaesthesia: a manikin study*
Authors:
- 7 December 2007
- journal article
- Published by Wiley in Anaesthesia
- Vol. 63 (1) , 65-70
Abstract
Summary: Direct laryngoscopy is performed regularly in paediatric anaesthetic practice for the purpose of intubating a patient's trachea. A minimum illumination of 700 lux at a distance of 20 mm has been suggested in a draft standard for laryngoscopes from the International Organization for Standardization. We investigated the minimum and optimum illumination required by anaesthetists during laryngoscopy. Fifty anaesthetists were recruited to perform laryngoscopy on a child manikin with a selection of laryngoscope blades and bulbs attached to a variable voltage supply. There was a large variation in the illumination range of different bulb and blade combinations. We demonstrated a large interindividual variability between anaesthetists for illumination requirements during laryngoscopy. Anaesthetists can see the larynx at very low light levels, but optimum illumination was significantly greater than minimum illumination. Two of the six combinations could produce more than 700 lux. In all, 78% and 54% of anaesthetists considered 700 lux too bright with a halogen and xenon bulb, respectively. Our study suggests that the proposed standard of 700 lux may possibly be too bright.Keywords
This publication has 10 references indexed in Scilit:
- Minimum and optimum light output of Macintosh size 3 laryngoscopy blades: a manikin study*Anaesthesia, 2007
- Flexibility and light emission of disposable paediatric Miller 1 laryngoscope blades*Anaesthesia, 2006
- The reliability of laryngoscope lightsAnaesthesia, 2006
- The Cardiff paediatric laryngoscope and its ‘non‐conforming’ siblingsAnaesthesia, 2005
- An assessment of the luminance and light field characteristics of used direct laryngoscopesCanadian Journal of Anesthesia/Journal canadien d'anesthésie, 1999
- A study of the brightness of laryngoscope lightAnaesthesia, 1996
- Light intensity and area of illumination provided by various laryngoscope bladesCanadian Journal of Anesthesia/Journal canadien d'anesthésie, 1994
- Tracheal intubation and laryngoscope designCanadian Journal of Anesthesia/Journal canadien d'anesthésie, 1993
- Laryngoscope design and the difficult adult tracheal intubationCanadian Journal of Anesthesia/Journal canadien d'anesthésie, 1989
- The Macintosh laryngoscopeAnaesthesia, 1984