Model Predictive Control of the Blood Glucose Concentration for Critically Ill Patients in Intensive Care Units
- 1 December 2019
- conference paper
- Published by Institute of Electrical and Electronics Engineers (IEEE)
- p. 3762-3769
Abstract
In this paper we present a linear model predictive control (MPC) algorithm for control of the blood glucose concentration of critically ill patients in an intensive care unit (ICU). We present a version of the algorithm that is suitable for automatic control. The algorithm can administer insulin and glucose. The insulin and glucose are administered using a parenteral (intravenous) route. We use a hysteresis logic to switch between insulin and glucose administration to prevent simultaneous administration of insulin and glucose. We test our algorithm in silico using three different simulation models to demonstrate the closed-loop performance of the algorithm. The MPC algorithm is based on linear models in the form of parsimonious second-order transfer functions describing the effect of intravenously injected insulin and glucose on the blood glucose concentration and a first-order disturbance transfer function containing an integrator to ensure offset-free control. The numerical results show that even in the ideal case without any measurement delay, tight glycemic control (80-110 mg/dL) cannot be achieved. Moreover, parenteral glucose is a requirement to avoid hypoglycemia and ensure extended tight glycemic control (60-140 mg/dL).Keywords
This publication has 29 references indexed in Scilit:
- Blood Glucose Control in the Intensive Care Unit: Benefits and RisksSeminars in Dialysis, 2010
- Intensive versus Conventional Glucose Control in Critically Ill PatientsNew England Journal of Medicine, 2009
- Health technology assessment review: Computerized glucose regulation in the intensive care unit - how to create artificial controlCritical Care, 2009
- A simulation model of glucose regulation in the critically illPhysiological Measurement, 2008
- Intensive Insulin Therapy in Mixed Medical/Surgical Intensive Care UnitsDiabetes, 2006
- ESPEN Guidelines on Enteral Nutrition: Intensive careClinical Nutrition, 2006
- Model-based glycaemic control in critical care—A review of the state of the possibleBiomedical Signal Processing and Control, 2006
- Model predictive glycaemic regulation in critical illness using insulin and nutrition input: A pilot studyMedical Engineering & Physics, 2005
- Intensive Insulin Therapy in Critically Ill PatientsNew England Journal of Medicine, 2001
- Physiologic evaluation of factors controlling glucose tolerance in man: measurement of insulin sensitivity and beta-cell glucose sensitivity from the response to intravenous glucose.Journal of Clinical Investigation, 1981