Development of high resolution programmable oversampling MI sensor system with 32-bit ADC for multi-channel bio-magnetic measurements
Authors:
Open Access
- 1 December 2019
- journal article
- conference paper
- Published by AIP Publishing in AIP Advances
- Vol. 9 (12) , 125243
Abstract
In order to further improving noise performance and achieving higher spatial resolution for bio-magnetic sensing, there is a tendency to develop a multi-channel integrated system for micro magnetic sensors. In this study, we have developed a high resolution programmable oversampling MI sensor system, as a unit module for multi-channel integrated bio-sensor system, based on Pk-pk VD-type MI magnetometer. We have achieved a high field detection sensitivity with good linearity, and a noise level lower than 1 pT/Hz1/2, by utilizing the time-differential measurement in each MI element for suppressing the low frequency noise components and increasing the stability of sensor system. Meanwhile, the proposed MI sensor system can be easily enlarged into multi-channel system, and is more suitable for integration and mass production. Furthermore, we have successfully achieved the real time R peak and T wave measurements of magnetocardiography (MCG) via new MI sensor system, with a high real time signal-to-noise ratio (SNR) of about 20 dB. We have at the first time achieved the clear P wave measurements of MCG using MI sensor, in an unshielded environment. The real time MCG measurements via new MI sensor system at room temperature, in an unshielded environment, will contribute to the diagnosis of heart disease.Keywords
Funding Information
- Japan Science and Technology Agency (AS2915024R)
This publication has 12 references indexed in Scilit:
- Alpha Rhythm and Visual Event-Related Fields Measurements at Room Temperature Using Magneto-Impedance Sensor SystemIEEE Transactions on Magnetics, 2019
- Development of Active Shielding-Type MI Gradiometer and Application for MagnetocardiographyIEEE Transactions on Magnetics, 2017
- High Performance Single Element MI Magnetometer With Peak-to-Peak Voltage Detector by Synchronized SwitchingIEEE Transactions on Magnetics, 2017
- Recent Advances of Amorphous Wire CMOS IC Magneto-Impedance Sensors: Innovative High-Performance Micromagnetic Sensor ChipJournal of Sensors, 2015
- Magnetocardiogram measured by fundamental mode orthogonal fluxgate arrayJournal of Applied Physics, 2015
- Highly sensitive magnetometer based on the off‐diagonal GMI effect in Co‐rich glass‐coated microwirePhysica Status Solidi (a), 2014
- Measurement of Spontaneous Oscillatory Magnetic Field of Guinea-Pig Smooth Muscle Preparation Using Pico-Tesla Resolution Amorphous Wire Magneto-Impedance SensorIEEE Transactions on Magnetics, 2011
- Magnetocardiography with sensors based on giant magnetoresistanceApplied Physics Letters, 2011
- Amorphous Wire and CMOS IC Based Magneto-Impedance Sensors—Origin, Topics, and FutureSensor Letters, 2007
- Off-Diagonal Impedance in Amorphous Wires and Its Application to Linear Magnetic SensorsIEEE Transactions on Magnetics, 2004