Open-source phantom with dedicated in-house software for image quality assurance in hybrid PET systems
Open Access
- 7 April 2025
- journal article
- research article
- Published by Springer Nature in EJNMMI Physics
- Vol. 12 (1) , 1-22
Abstract
Patients’ diagnosis, treatment and follow-up increasingly rely on multimodality imaging. One of the main limitations for the optimal implementation of hybrid systems in clinical practice is the time and expertise required for applying standardized protocols for equipment quality assurance (QA). Experimental phantoms are commonly used for this purpose, but they are often limited to a single modality and single quality parameter, lacking automated analysis capabilities. In this study, we developed a multimodal 3D-printed phantom and software for QA in positron emission tomography (PET) hybrid systems, with computed tomography (CT) or magnetic resonance (MR), by assessing signal, spatial resolution, radiomic features, co-registration and geometric distortions. Phantom models and Python software for the proposed QA are available to download, and a user-friendly plugin compatible with the open-source 3D-Slicer software has been developed. The QA viability was proved by characterizing a Philips-Gemini-TF64-PET/CT in terms of signal response (mean, µ), intrinsic variability for three consecutive measurements (daily variation coefficient, CoVd) and reproducibility over time (variation coefficient across 5 months, CoVm). For this system, averaged recovery coefficient for activity concentration was µ = 0.90 ± 0.08 (CoVd = 0.6%, CoVm = 9%) in volumes ranging from 7 to 42 ml. CT calibration-curve averaged over time was $$\text{HU}=(951\pm 12)\times \text{density}-(944\pm 15)$$ with variability of slope and y-intercept of (CoVd = 0.4%, CoVm = 1.2%) and (CoVd = 0.4%, CoVm = 1.6%), respectively. Radiomics reproducibility resulted in (CoVd = 18%, CoVm = 30%) for PET and (CoVd = 15%, CoVm = 22%) for CT. Co-registration was assessed by Dice-Similarity-Coefficient (DSC) along 37.8 cm in superior-inferior (z) direction (well registered if DSC ≥ 0.91 and Δz ≤ 2 mm), resulting in 3/7 days well co-registered. Applicability to other scanners was additionally proved with Philips-Vereos-PET/CT (V), Siemens-Biograph-Vison-600-PET/CT (S) and GE-SIGNA-PET/MR (G). PET concentration accuracy was (µ = 0.86, CoVd = 0.3%) for V, (µ = 0.87, CoVd = 0.8%) for S, and (µ = 1.10, CoVd = 0.34%) for G. MR(T2) was well co-registered with PET in 3/4 cases, did not show significant distortion within a transaxial diameter of 27.8 cm and along 37 cm in z, and its radiomic variability was CoVd = 13%. Open-source QA protocol for PET hybrid systems has been presented and its general applicability has been proved. This package facilitates simultaneously simple and semi-automated evaluation for various imaging modalities, providing a complete and efficient QA solution.
Keywords
Funding Information
- Instituto de Salud Carlos III (FI23/00097, CP23/000106)
This publication has 56 references indexed in Scilit:
- A multi‐modality physical phantom for mimicking tumor heterogeneity patterns in PET/CT and PET/MRIMedical Physics, 2022
- Advanced quantitative evaluation of PET systems using the ACR phantom and NiftyPET softwareMedical Physics, 2022
- Design and evaluation of a modular multimodality imaging phantom to simulate heterogeneous uptake and enhancement patterns for radiomic quantification in hybrid imaging: A feasibility studyMedical Physics, 2022
- PET/CT acceptance testing and quality assurance: Executive summary of AAPM Task Group 126 ReportMedical Physics, 2020
- Assessment of PET and SPECT phantom image quality through automated binary classification of cold rod arraysMedical Physics, 2019
- Repeatability of 18F‐FDG PET radiomic features: A phantom study to explore sensitivity to image reconstruction settings, noise, and delineation methodMedical Physics, 2018
- Recent advances on the development of phantoms using 3D printing for imaging with CT, MRI, PET, SPECT, and ultrasoundMedical Physics, 2018
- 3D printed phantoms mimicking cortical bone for the assessment of ultrashort echo time magnetic resonance imagingMedical Physics, 2017
- Use of image registration and fusion algorithms and techniques in radiotherapy: Report of the AAPM Radiation Therapy Committee Task Group No. 132Medical Physics, 2017
- PET/MRI: Where might it replace PET/CT?Journal of Magnetic Resonance Imaging, 2017