Different approaches for quantifying ventilation distribution and lung tissue properties by functional EIT
Authors:
- G HahnGHG Hahn
- J DittmarJDJ Dittmar
- A JustAJA Just
- M QuintelMQM Quintel
- G HelligeGHG Hellige
- 21 July 2010
- journal article
- Published by IOP Publishing in Physiological Measurement
- Vol. 31 (8) , S73-S84
Abstract
We investigated five different methods which can be applied to quantitatively construct functional tomograms of the lungs. The focus was on the sensitivity of functional tomograms to errors in acquired data. To quantify this sensitivity, theoretical, error-free data sets of well-known properties were artificially generated based on a 'living thorax model'. Physiological time courses and a typical distribution of errors caused by a typical Goe-MF II EIT system were used for the calculations which encompassed a range up to 50 times greater than the initial error level (4 microV(rms max)-400 microV(rms max)). Additionally, low-pass filtering and principal component analysis (PCA) were used to quantify the effect of preprocessing the raw data. The results demonstrate that all methods based on fitting the local to the global time course were superior to the common functional tomograms utilizing standard deviation or maximum and minimum detection. Ventilation distribution was best quantified by the so-called VT methods. Filling capacity--a lung tissue property--was least dependent on increasing error levels. The errors introduced by filtering are significant with respect to a quantitative analysis of ventilation distribution. A preprocessing of raw data by applying a PCA performed well on the data sets which had been constructed but were, nonetheless, realistic. This approach appears to be highly promising for application on real data which is known to be erroneous.Keywords
This publication has 26 references indexed in Scilit:
- A comparison of two EIT systems suitable for imaging impedance changes in epilepsyPhysiological Measurement, 2009
- Dynamic separation of pulmonary and cardiac changes in electrical impedance tomographyPhysiological Measurement, 2008
- Analysis of resting noise characteristics of three EIT systems in order to compare suitability for time difference imaging with scalp electrodes during epileptic seizuresPhysiological Measurement, 2007
- Lung Volume Recruitment after Surfactant Administration Modifies Spatial Distribution of VentilationAmerican Journal of Respiratory and Critical Care Medicine, 2006
- Electrical impedance tomography in extremely prematurely born infants and during high frequency oscillatory ventilation analyzed in the frequency domainPhysiological Measurement, 2006
- Propagation of measurement noise through backprojection reconstruction in electrical impedance tomographyIEEE Transactions on Medical Imaging, 2002
- Detection of local lung air content by electrical impedance tomography compared with electron beam CTJournal of Applied Physiology, 2002
- Gravity effects on regional lung ventilation determined by functional EIT during parabolic flightsJournal of Applied Physiology, 2001
- Quantification in impedance imagingClinical Physics and Physiological Measurement, 1990
- RESISTIVITY MODELLING FOR ARBITRARILY SHAPED TWO‐DIMENSIONAL STRUCTURES*Geophysical Prospecting, 1979