Analysis of resting noise characteristics of three EIT systems in order to compare suitability for time difference imaging with scalp electrodes during epileptic seizures
Authors:
- 26 June 2007
- journal article
- Published by IOP Publishing in Physiological Measurement
- Vol. 28 (7) , S217-S236
Abstract
Electrical impedance tomography measurements in clinical applications are limited by an undesired noise component. We have investigated the noise in three systems suitable for imaging epileptic seizures, the UCH Mark1b, UCH Mark2.5 and KHU Mark1 16 channel, at applied frequencies in three steps from 1 to 100 kHz, by varying load impedance, single terminal or multiplexed measurements, and in test objects of increasing complexity from a resistor to a saline filled tank and human volunteer. The noise was white, and increased from about 0.03% rms on the resistor to 0.08% on the human; it increased with load but was independent of use of the multiplexer. The KHU Mark1 delivered the best performance with noise spectra of about 0.02%, which could be further reduced by averaging to a level where reliable imaging of changes of about 0.1% estimated during epileptic seizures appears plausible.Keywords
This publication has 24 references indexed in Scilit:
- Factors limiting the application of electrical impedance tomography for identification of regional conductivity changes using scalp electrodes during epileptic seizures in humansPhysiological Measurement, 2006
- Tissue Resistance Changes and the Profile of Synchronized Neuronal Activity During Ictal Events in the Low-Calcium Model of EpilepsyJournal of Neurophysiology, 2004
- Propagation of measurement noise through backprojection reconstruction in electrical impedance tomographyIEEE Transactions on Medical Imaging, 2002
- Quantitative Evaluation of the Performance of Different Electrical Tomography Devices - Quantitative Evaluierung der Systemeigenschaften verschiedener Elektrotomographie-GeräteBiomedical Engineering / Biomedizinische Technik, 2001
- Electrical impedance tomography (EIT) in applications related to lung and ventilation: a review of experimental and clinical activitiesPhysiological Measurement, 2000
- A simple method to check the dynamic performance of electrical impedance tomography systemsPhysiological Measurement, 2000
- A Cole phantom for EITPhysiological Measurement, 1995
- The Sheffield data collection systemClinical Physics and Physiological Measurement, 1987
- Impedance Changes during Epileptic SeizuresEpilepsia, 1966
- Spectra of Quantized SignalsBell System Technical Journal, 1948