Abstract
We describe an optical setup that performs optical coherence tomography (OCT) and fluorescence microscopy on the mouse retina through a shared sample arm. The crystalline lens of the animal’s eye is the objective lens for both, and a tube-lens and scan-lens relay in a modified 4 f configuration works around the small entrance pupil of the eye while leaving room to mount a dichroic mirror. The platform reaches a measured axial resolution of 3.33 µm in air across a 1.4 mm field of view and 3.9 µm lateral resolution in the fluorescence path. We report how the collimator was chosen, how the relay was designed, which aberrations dominate, and how we reduced them.