Body fluid cellular analysis using the Sysmex XN‐2000 automatic hematology analyzer: focusing on malignant samples
Authors:
- 12 September 2014
- journal article
- research article
- Published by Wiley in International Journal of Laboratory Hematology
- Vol. 37 (3) , 346-356
Abstract
Summary: Introduction: The majority of previous studies on body fluid (BF) mode of automatic hematology analyzer used nonmalignant BF samples. Here, we evaluated the BF mode on the recently launched Sysmex XN for counting blood cells, especially for malignant samples.Methods: A total of 405 BF specimens including 125 malignant samples were analyzed using both the automated method and manual microscopy.Results: In non‐cerebrospinal fluids (CSF) samples, there was an agreement between two methods for WBC, RBC, polymorphonuclear, and mononuclear cell counts (R2 = 0.96, 0.94, 0.88, and 0.88, respectively). CSF samples showed slightly poorer correlations than other fluids. Exclusion of malignant samples significantly improved correlations in non‐CSF samples, but not in CSF samples. High fluorescence‐BF (HF‐BF) cells were identified significantly more frequently in malignant samples compared to benign samples (17.8 and 4.15/100 WBC, respectively; P < 0.001). Receiver operating characteristic curve analysis demonstrated an HF‐BF cell AUC of 0.791 using a cutoff value of 6.9/100 WBC for detecting malignant samples.Conclusion: The BF mode on the Sysmex XN could be an alternative method for the manual counts in the BF analysis with a few drawbacks. However, if a concentration of HF‐BF cells is greater than the given threshold, microscopic examination should be subsequently performed.Keywords
This publication has 15 references indexed in Scilit:
- Cellular origin and diagnostic significance of high‐fluorescent cells in cerebrospinal fluid detected by the XE‐5000 hematology analyzerInternational Journal of Laboratory Hematology, 2013
- Analysis of serous body fluids using the CELL-DYN Sapphire hematology analyzerClinical Chemistry and Laboratory Medicine (CCLM), 2012
- Performance Evaluation of the Sysmex XE-5000 Hematology Analyzer for White Blood Cell Analysis in Cerebrospinal FluidArchives of Pathology & Laboratory Medicine, 2012
- Automated vs. manual cerebrospinal fluid cell counts: a work and cost analysis comparing the Sysmex XE-5000 and the Fuchs-Rosenthal manual counting chamberInternational Journal of Laboratory Hematology, 2011
- Automated Cerebrospinal Fluid Cell Counts Using the Sysmex XE-5000American Journal of Clinical Pathology, 2010
- ORIGINAL ARTICLE: Performance evaluation of the body fluid mode on the platform Sysmex XE‐5000 series automated hematology analyzerInternational Journal of Laboratory Hematology, 2010
- Use of the Cell-Dyn Sapphire Hematology Analyzer for Automated Counting of Blood Cells in Body FluidsAmerican Journal of Clinical Pathology, 2010
- Evaluation of the XE-5000 for the automated analysis of blood cells in cerebrospinal fluidClinical Biochemistry, 2009
- Identification and quantification of high fluorescence‐stained lymphocytes as antibody synthesizing/secreting cells using the automated routine hematology analyzer XE‐2100Cytometry Part B: Clinical Cytometry, 2007
- Automated Flow Cytometric Analysis of Blood Cells in Cerebrospinal FluidAmerican Journal of Clinical Pathology, 2004