Comparison of low contrast detectability between a digital amorphous silicon and a screen‐film based imaging system for thoracic radiography
- 12 July 1999
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 26 (7) , 1349-1358
Abstract
Low contrast threshold detectability is investigated theoretically and experimentally for an amorphous silicon x‐ray detector designed for digital radiography and for a standard thoracic screen‐film combination. A theoretical signal‐to‐noise ratio is described with a human observer signal detection model. It is calculated using the detective quantum efficiency (DQE) and the modulation transfer function of the imaging system, as well as a spatial response function for the human visual system. Using a four‐alternative forced choice observer perception study, the threshold contrasts of disk shaped objects ranging in size from 0.3 to 4.0 mm are determined. Significantly better contrast detectability is obtained from the digital detector, which is attributed to its higher DQE. On average, the disk shaped objects can be detected at 45% less contrast than required for screen‐film. With no free parameters the experimental data agree well with those predicted by the observer model. Based upon the data, the model predicts that x‐ray exposure for the detector only needs to be 30% of the exposure required to perform equally well on the contrast‐detail detectability task using screen‐film.Keywords
This publication has 40 references indexed in Scilit:
- Detection of Simulated Chest Lesions with Normal and Reduced Radiation DoseInvestigative Radiology, 1998
- Perceived noise versus display noise in temporally filtered image sequencesJournal of Electronic Imaging, 1996
- Evaluation of an asymmetric screen‐film system for chest radiographyMedical Physics, 1994
- A Fully Automated Adaptive Unsharp Masking Technique in Digital Chest RadiographInvestigative Radiology, 1992
- Effect of pixel size on detectability of low-contrast signals in digital radiographyJournal of the Optical Society of America A, 1987
- Histogram-Directed Processing of Digital Chest ImagesInvestigative Radiology, 1986
- Digital imaging of the chestJournal of Thoracic Imaging, 1985
- Investigation of basic imaging properties in digital radiography. 4. Effect of unsharp masking on the detectability of simple patternsMedical Physics, 1985
- Investigation of basic imaging properties in digital radiography. 3. Effect of pixel size on SNR and threshold contrastMedical Physics, 1985
- Spatial and Temporal Contrast-Sensitivity Functions of the Visual SystemJournal of the Optical Society of America, 1966